Image not available in RecallScope.
Quadrox-i adult oxygenator w integrated hardshell reservoir
Official notice
30 Aug 2024
An official Health Canada notice concerns Quadrox-i adult oxygenator w integrated hardshell reservoir. Maquet Cardiopulmonary's monthly inspection of oxygenators for the presence of the Softline coating revealed that two out of ten oxygenators tested did not pass. The Softline coating is a vital component in the biocompatibility of blood contact components used in extracorporeal circulation. This means that the oxygenators installed in the products listed above do not have the Softline coating. The issue is limited to one lot number in Canada. Recall Start Date: August 20, 2024 Published 2024-08-30.
What to do now
Contact the manufacturer if you require additional information.
Recall number
75990
Official notice
2024-08-30
The hazard
The hazard
Maquet Cardiopulmonary's monthly inspection of oxygenators for the presence of the Softline coating revealed that two out of ten oxygenators tested did not pass. The Softline coating is a vital component in the biocompatibility of blood contact components used in extracorporeal circulation. This means that the oxygenators installed in the products listed above do not have the Softline coating. The issue is limited to one lot number in Canada. Recall Start Date: August 20, 2024
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Manufacturer
Phone
Brand: Maquet Cardiopulmonary, Quadrox-i Adult Oxygenator W Integrated Hardshell Reservoir.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.