No image published for this recall

No image published for this recall

Maquet hemoconcentrator set

Official notice

18 Jan 2024

Last checked

18 Jan 2024

Medium riskHealth Canada

An official Health Canada notice concerns Maquet hemoconcentrator set. The component "hemoconcentrator" was assembled into the finished product (hemoconcentrator tubing set) after its shelf-life had expired. After internal testing, it was determined that only one component lot was affected. Therefore, this field action is limited to products that containing hemoconcentrator of the one affected batch. Recall start date: December 29, 2023 Published 2024-01-18.

What to do now

Contact the manufacturer if you require additional information.

Recall number

74926

Official notice

2024-01-18

The hazard

The component "hemoconcentrator" was assembled into the finished product (hemoconcentrator tubing set) after its shelf-life had expired. After internal testing, it was determined that only one component lot was affected. Therefore, this field action is limited to products that containing hemoconcentrator of the one affected batch. Recall start date: December 29, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Lots

was

Recall Numbers

74926

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Hemoconcentrator Set With Softline Coating, Lot or serial number - 3000277066 3000276250, Model or catalog number - BO-H 48771

Product 2

Affected products - Hemoconcentrator Set, Lot or serial number - 3000273187, Model or catalog number - H 52570

Risk / hazard

The component "hemoconcentrator" was assembled into the finished product (hemoconcentrator tubing set) after its shelf-life had expired. After internal testing, it was determined that only one component lot was affected. Therefore, this field action is limited to products that containing hemoconcentrator of the one affected batch. Recall start date: December 29, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Maquet Cardiopulmonary, Maquet Hemoconcentrator Set.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

74926

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 18 Jan 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.