Heater-cooler unit hcu 40

Official notice

6 Feb 2025

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Heater-cooler unit hcu 40. The ice bracket is attached to the inlet of the cardioplegia tank where ice is produced. To avoid interrupting the water circulation on the cardioplegia side, ice must not cover the opening at the bottom of the tank. Due to its very high thermal conductivity, the ice bracket prevents ice buildup in the affected area and the opening always remains patent. An internal investigation came to the conclusion that the failure rate of the ice bracket becomes unacceptably high after two years. For devices where the ice bracket was replaced more than 2 years ago: The ice bracket shall be replaced as soon as possible free of charge. For devices where the ice bracket was replaced less than 2 years ago: The ice bracket shall be replaced during the next preventive maintenance free of charge. In the future (not part of this field action): The ice bracket will be replaced biennially during preventive maintenance. Recall start date: January 24, 2025 Published 2025-02-06.

What to do now

Contact the manufacturer if you require additional information.

Recall number

76912

Official notice

2025-02-06

The hazard

The ice bracket is attached to the inlet of the cardioplegia tank where ice is produced. To avoid interrupting the water circulation on the cardioplegia side, ice must not cover the opening at the bottom of the tank. Due to its very high thermal conductivity, the ice bracket prevents ice buildup in the affected area and the opening always remains patent. An internal investigation came to the conclusion that the failure rate of the ice bracket becomes unacceptably high after two years. For devices where the ice bracket was replaced more than 2 years ago: The ice bracket shall be replaced as soon as possible free of charge. For devices where the ice bracket was replaced less than 2 years ago: The ice bracket shall be replaced during the next preventive maintenance free of charge. In the future (not part of this field action): The ice bracket will be replaced biennially during preventive maintenance. Recall start date: January 24, 2025

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

76912

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - Heater-Cooler Unit HCU 40, Lot or serial numver - More than 10 numbers, contact manufacturer., Model or catalogue number - 70104.4054 (200-240V) 70105.4917 (110-120V)

Hazard

The ice bracket is attached to the inlet of the cardioplegia tank where ice is produced. To avoid interrupting the water circulation on the cardioplegia side, ice must not cover the opening at the bottom of the tank. Due to its very high thermal conductivity, the ice bracket prevents ice buildup in the affected area and the opening always remains patent. An internal investigation came to the conclusion that the failure rate of the ice bracket becomes unacceptably high after two years. For devices where the ice bracket was replaced more than 2 years ago: The ice bracket shall be replaced as soon as possible free of charge. For devices where the ice bracket was replaced less than 2 years ago: The ice bracket shall be replaced during the next preventive maintenance free of charge. In the future (not part of this field action): The ice bracket will be replaced biennially during preventive maintenance. Recall start date: January 24, 2025

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Maquet Cardiopulmonary, Heater-cooler Unit Hcu 40.

Source & status

Tier

official

Category

Health

Official classification

Type II

Recall details

Recall number

76912

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 6 Feb 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button