Open official notice

Maquet hemoconcentrator set with softline coating

Official notice

18 Dec 2024

RecallScope risk estimate:CriticalHealth Canada

An official Health Canada notice concerns Maquet hemoconcentrator set with softline coating. An issue with eight technical drawings, specifically the absence of the bonding symbol for certain connections on the tubing set was found. It is important to note that the specified bonding procedure, which involved the use of cyclohexanone, was not carried out. While the absence of the bonding on the connectors may not be immediately apparent, it could potentially lead to leakage in the affected connections. Recall start date: December 9. 2024 Published 2024-12-18.

What to do now

Contact the manufacturer if you require additional information.

Recall number

76715

Official notice

2024-12-18

The hazard

An issue with eight technical drawings, specifically the absence of the bonding symbol for certain connections on the tubing set was found. It is important to note that the specified bonding procedure, which involved the use of cyclohexanone, was not carried out. While the absence of the bonding on the connectors may not be immediately apparent, it could potentially lead to leakage in the affected connections. Recall start date: December 9. 2024

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

76715

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - Hemoconcentrator Set With Softline Coating, Lot or serial number - 3000382838 701052273 3000361271 701074294 3000407060 701035333 3000393741, Model or catalog number - BO-H 48773

Hazard

An issue with eight technical drawings, specifically the absence of the bonding symbol for certain connections on the tubing set was found. It is important to note that the specified bonding procedure, which involved the use of cyclohexanone, was not carried out. While the absence of the bonding on the connectors may not be immediately apparent, it could potentially lead to leakage in the affected connections. Recall start date: December 9. 2024

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Maquet Cardiopulmonary, Hemoconcentrator Set With Softline Coating.

Source & status

Tier

official

Category

Health

Official classification

Type II

Recall details

Recall number

76715

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 18 Dec 2024. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.