Image not available in RecallScope.
Bs 3 8x3 32 l1.7 (bubble sensor for 3 8” x 3 32” tubing, length 1,7 m)
Official notice
7 Jan 2026
An official Health Canada notice concerns Bs 3 8x3 32 l1.7 (bubble sensor for 3 8” x 3 32” tubing, length 1,7 m). Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Bubble Sensor. Excessive bending of the cable can lead to poor contact, which may trigger errors on the connected medical device. These errors can occur temporarily when the cable is moved or permanently if the connection is fully compromised. Published 2026-01-07.
What to do now
Contact the manufacturer if you require additional information.
Recall number
81458
Official notice
2026-01-07
The hazard
The hazard
Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Bubble Sensor. Excessive bending of the cable can lead to poor contact, which may trigger errors on the connected medical device. These errors can occur temporarily when the cable is moved or permanently if the connection is fully compromised.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Other
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Maquet Cardiopulmonary.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.