Open official notice

Rotaflow console

Official notice

29 Oct 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Rotaflow console. During internal validation of new replacement batteries, MCP identified an issue on the RotaFlow console. This issue can cause the device to stop battery charging after the battery operation mode has been utilized. Further investigation identified a software problem as root cause. Because of this, all devices on the market are considered affected by this issue. Published 2026-10-29.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82878

Official notice

2026-10-29

The hazard

During internal validation of new replacement batteries, MCP identified an issue on the RotaFlow console. This issue can cause the device to stop battery charging after the battery operation mode has been utilized. Further investigation identified a software problem as root cause. Because of this, all devices on the market are considered affected by this issue.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

RA RA-82878

Recall Numbers

82878

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 70104.3292

Hazard

During internal validation of new replacement batteries, MCP identified an issue on the RotaFlow console. This issue can cause the device to stop battery charging after the battery operation mode has been utilized. Further investigation identified a software problem as root cause. Because of this, all devices on the market are considered affected by this issue.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Maquet Cardiopulmonary.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82878

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 29 Oct 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button