Ctp - tubing sets

Official notice

15 Jul 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Ctp - tubing sets. The manufacturer became aware of this issue due to six customer complaint reports. It was found that the sterile barrier was compromised, as the Tyvek lid was found opened prior to use. No harm to any person was reported. An internal investigation has confirmed that all affected products show a consistent similar defect pattern occurring at the same sealing area and were manufactured using the same tool. Therefore, this FSCA is limited to all sterile products within shelf-life manufactured using the same tool. Consignees will be asked to confirm receipt of the field action notification. Consignees will be asked to return all unused, affected sold product from consignee locations. Published 2026-07-15.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82328

Official notice

2026-07-15

The hazard

The manufacturer became aware of this issue due to six customer complaint reports. It was found that the sterile barrier was compromised, as the Tyvek lid was found opened prior to use. No harm to any person was reported. An internal investigation has confirmed that all affected products show a consistent similar defect pattern occurring at the same sealing area and were manufactured using the same tool. Therefore, this FSCA is limited to all sterile products within shelf-life manufactured using the same tool. Consignees will be asked to confirm receipt of the field action notification. Consignees will be asked to return all unused, affected sold product from consignee locations.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

3000419748, 3000407529, 3000415791, 3000460078, 3000491063, 3000468700, 3000460082, 3000407530, 3000491064

Skus

BO-HQV 59206, BEQ-HQV 52700-1, BEQ-HQV 46904, BO-HQV 55004

Other

RA RA-82328

Model numbers

BO-HQV 59206, BEQ-HQV 52700-1, BEQ-HQV 46904, BO-HQV 55004

Recall Numbers

82328

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - 3000419748, Model or catalogue number - BO-HQV 59206

Product 2

Lot or serial number - 3000407529, Model or catalogue number - BEQ-HQV 52700-1

Product 3

Lot or serial number - 3000415791, Model or catalogue number - BEQ-HQV 52700-1

Product 4

Lot or serial number - 3000460078, Model or catalogue number - BEQ-HQV 52700-1

Product 5

Lot or serial number - 3000491063, Model or catalogue number - BEQ-HQV 52700-1

Product 6

Lot or serial number - 3000468700, Model or catalogue number - BEQ-HQV 52700-1

Product 7

Lot or serial number - 3000460082, Model or catalogue number - BEQ-HQV 52700-1

Product 8

Lot or serial number - 3000407530, Model or catalogue number - BEQ-HQV 52700-1

Product 9

Lot or serial number - 3000491064, Model or catalogue number - BEQ-HQV 52700-1

Product 10

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - BEQ-HQV 46904

Product 11

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - BO-HQV 55004

Hazard

The manufacturer became aware of this issue due to six customer complaint reports. It was found that the sterile barrier was compromised, as the Tyvek lid was found opened prior to use. No harm to any person was reported. An internal investigation has confirmed that all affected products show a consistent similar defect pattern occurring at the same sealing area and were manufactured using the same tool. Therefore, this FSCA is limited to all sterile products within shelf-life manufactured using the same tool. Consignees will be asked to confirm receipt of the field action notification. Consignees will be asked to return all unused, affected sold product from consignee locations.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Maquet Cardiopulmonary.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82328

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 15 Jul 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button