Astral 150™

Official notice

9 Jul 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Astral 150™. An internal electrical component (supercapacitor) may leak electrolyte over time. In rare cases, this may damage specific circuitry on the ventilator's Printed Circuit Board Assembly ("PCBA"), causing the ventilator to inadvertently enter a fail-safe state. In this state, the ventilator is no longer able to deliver therapy. Alternative means of ventilation must be provided. Published 2026-07-09.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82300

Official notice

2026-07-09

The hazard

An internal electrical component (supercapacitor) may leak electrolyte over time. In rare cases, this may damage specific circuitry on the ventilator's Printed Circuit Board Assembly ("PCBA"), causing the ventilator to inadvertently enter a fail-safe state. In this state, the ventilator is no longer able to deliver therapy. Alternative means of ventilation must be provided.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

27073, 27003

Other

RA RA-82300

Model numbers

27073, 27003

Recall Numbers

82300

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 27073

Product 2

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 27003

Hazard

An internal electrical component (supercapacitor) may leak electrolyte over time. In rare cases, this may damage specific circuitry on the ventilator's Printed Circuit Board Assembly ("PCBA"), causing the ventilator to inadvertently enter a fail-safe state. In this state, the ventilator is no longer able to deliver therapy. Alternative means of ventilation must be provided.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Manufacturer

Not provided

Phone

Not provided

Brand: Resmed.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82300

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 9 Jul 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button