Rapicide™ pa high level disinfectant
Official notice
17 Sept 2026
Last checked
13 Nov 2026
An official Health Canada notice concerns Rapicide™ pa high level disinfectant. STERIS identified through customer complaints that a small number of RAPICIDE™ Part A bottles within certain lots may begin to bulge or expand over time. Part A is the active component of chemistry that creates peracetic acid, which may off-gas during the shelf-life of the product. Under routine conditions, the Part A bottle cap includes a vent intended to allow gas to escape (off-gassing) and reduce the potential for gas accumulation. If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.STERIS will provide a customer notification letter to all consignees who purchased affected lots. Consignees will be requested to follow the appropriate procedures and PPE needed to address any swollen bottles. The chemistry itself remains safe and effective to use in accordance with its labeling. However, replacement product will be provided for any product that had to be disposed. Published 2026-09-17.
What to do now
Contact the manufacturer if you require additional information.
Recall number
82643
Official notice
2026-09-17
The hazard
The hazard
STERIS identified through customer complaints that a small number of RAPICIDE™ Part A bottles within certain lots may begin to bulge or expand over time. Part A is the active component of chemistry that creates peracetic acid, which may off-gas during the shelf-life of the product. Under routine conditions, the Part A bottle cap includes a vent intended to allow gas to escape (off-gassing) and reduce the potential for gas accumulation. If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.STERIS will provide a customer notification letter to all consignees who purchased affected lots. Consignees will be requested to follow the appropriate procedures and PPE needed to address any swollen bottles. The chemistry itself remains safe and effective to use in accordance with its labeling. However, replacement product will be provided for any product that had to be disposed.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Lots
Skus
Other
Model numbers
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Product 2
Product 3
Product 4
Product 5
Product 6
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Medivators.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.