Rapicide™ pa high level disinfectant

Official notice

17 Sept 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Rapicide™ pa high level disinfectant. STERIS identified through customer complaints that a small number of RAPICIDE™ Part A bottles within certain lots may begin to bulge or expand over time. Part A is the active component of chemistry that creates peracetic acid, which may off-gas during the shelf-life of the product. Under routine conditions, the Part A bottle cap includes a vent intended to allow gas to escape (off-gassing) and reduce the potential for gas accumulation. If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.STERIS will provide a customer notification letter to all consignees who purchased affected lots. Consignees will be requested to follow the appropriate procedures and PPE needed to address any swollen bottles. The chemistry itself remains safe and effective to use in accordance with its labeling. However, replacement product will be provided for any product that had to be disposed. Published 2026-09-17.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82643

Official notice

2026-09-17

The hazard

STERIS identified through customer complaints that a small number of RAPICIDE™ Part A bottles within certain lots may begin to bulge or expand over time. Part A is the active component of chemistry that creates peracetic acid, which may off-gas during the shelf-life of the product. Under routine conditions, the Part A bottle cap includes a vent intended to allow gas to escape (off-gassing) and reduce the potential for gas accumulation. If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.STERIS will provide a customer notification letter to all consignees who purchased affected lots. Consignees will be requested to follow the appropriate procedures and PPE needed to address any swollen bottles. The chemistry itself remains safe and effective to use in accordance with its labeling. However, replacement product will be provided for any product that had to be disposed.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

689982, 690994, 689172, 692571, 691740, 694358

Skus

ML02-0116

Other

RA RA-82643

Model numbers

ML02-0116

Recall Numbers

82643

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - 689982, Model or catalogue number - ML02-0116

Product 2

Lot or serial number - 690994, Model or catalogue number - ML02-0116

Product 3

Lot or serial number - 689172, Model or catalogue number - ML02-0116

Product 4

Lot or serial number - 692571, Model or catalogue number - ML02-0116

Product 5

Lot or serial number - 691740, Model or catalogue number - ML02-0116

Product 6

Lot or serial number - 694358, Model or catalogue number - ML02-0116

Hazard

STERIS identified through customer complaints that a small number of RAPICIDE™ Part A bottles within certain lots may begin to bulge or expand over time. Part A is the active component of chemistry that creates peracetic acid, which may off-gas during the shelf-life of the product. Under routine conditions, the Part A bottle cap includes a vent intended to allow gas to escape (off-gassing) and reduce the potential for gas accumulation. If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.STERIS will provide a customer notification letter to all consignees who purchased affected lots. Consignees will be requested to follow the appropriate procedures and PPE needed to address any swollen bottles. The chemistry itself remains safe and effective to use in accordance with its labeling. However, replacement product will be provided for any product that had to be disposed.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Medivators.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82643

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 17 Sept 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button