Open official notice

Cardiosave® hybrid and cardiosave® rescue intra-aortic balloon pumps

Official notice

19 Nov 2026

Last checked

22 Nov 2026

RecallScope risk estimate:HighHealth Canada

An official Health Canada notice concerns Cardiosave® hybrid and cardiosave® rescue intra-aortic balloon pumps. Getinge has identified that some CARDIOSAVE® devices may contain a power management board (PMB) and/or solenoid control board (SCB) with one or more capacitors installed with reversed polarity. These capacitors may degrade over time, potentially causing CARDIOSAVE® device battery charging failures, an inability to power on, interruption of therapy, or unexpected shutdown. The device may continue to function as intended, since the component degradation is not an immediate effect. Continued operation in this condition may eventually lead to a circuit malfunction, affecting critical system functions as outlined in the factors contributing to risk. Functional testing of the PMBs and SCBs may not identify every affected assembly unless capacitor polarity is specifically inspected and verified. Published 2026-11-19.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82982

Official notice

2026-11-19

The hazard

Getinge has identified that some CARDIOSAVE® devices may contain a power management board (PMB) and/or solenoid control board (SCB) with one or more capacitors installed with reversed polarity. These capacitors may degrade over time, potentially causing CARDIOSAVE® device battery charging failures, an inability to power on, interruption of therapy, or unexpected shutdown. The device may continue to function as intended, since the component degradation is not an immediate effect. Continued operation in this condition may eventually lead to a circuit malfunction, affecting critical system functions as outlined in the factors contributing to risk. Functional testing of the PMBs and SCBs may not identify every affected assembly unless capacitor polarity is specifically inspected and verified.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

0998-00-0800-53, 0998-00-0800-65

Other

RA RA-82982

Model numbers

0998-00-0800-53, 0998-00-0800-65

Recall Numbers

82982

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0998-00-0800-53

Product 2

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0998-00-0800-65

Hazard

Getinge has identified that some CARDIOSAVE® devices may contain a power management board (PMB) and/or solenoid control board (SCB) with one or more capacitors installed with reversed polarity. These capacitors may degrade over time, potentially causing CARDIOSAVE® device battery charging failures, an inability to power on, interruption of therapy, or unexpected shutdown. The device may continue to function as intended, since the component degradation is not an immediate effect. Continued operation in this condition may eventually lead to a circuit malfunction, affecting critical system functions as outlined in the factors contributing to risk. Functional testing of the PMBs and SCBs may not identify every affected assembly unless capacitor polarity is specifically inspected and verified.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Datascope.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82982

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 19 Nov 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button