Azurion 7 m20
Official notice
13 Nov 2026
Last checked
19 Nov 2026
An official Health Canada notice concerns Azurion 7 m20. Philips has identified two software issues affecting Azurion R3.1 systems that may result in loss of imaging (X-ray) functionality, and/or loss of motorized movement. Issue 1 - System remains in continuous restart mode after the start-up. Due to a software issue, the License Manager - which verifies installed licenses and enables optional functionality in the Azurion system based on the available licenses - expands the Windows registry database at every system start or restart. Over time, the registry may grow beyond the maximum size that the License Manager can load. When this happens, the License Manager cannot start and the system will enter a continuous reboot loop. The Azurion system will not complete booting, and the message "X-ray system is booting..." remains displayed on the screen.Issue 2 - Longitudinal position error - Applicable only to azurion systems with Poly-G3 frontal stand. When requesting a movement of the frontal stand in longitudinal direction, the Azurion system may detect a mismatch between the expected longitudinal (set-) position and the measured (actual) position of the stand. This mismatch is caused by a glitch in the longitudinal position value provided by the position sensing potentiometer.When the mismatch is detected, the motion power to the frontal stand is automatically cut off. The message "Some stand movements are not available" is displayed to the user.In this situation, X-ray imaging and table movements will continue to function.This issue can be resolved with a cold restart of the system or by restarting the geometry. Published 2026-11-13.
What to do now
Contact the manufacturer if you require additional information.
Recall number
82951
Official notice
2026-11-13
The hazard
The hazard
Philips has identified two software issues affecting Azurion R3.1 systems that may result in loss of imaging (X-ray) functionality, and/or loss of motorized movement. Issue 1 - System remains in continuous restart mode after the start-up. Due to a software issue, the License Manager - which verifies installed licenses and enables optional functionality in the Azurion system based on the available licenses - expands the Windows registry database at every system start or restart. Over time, the registry may grow beyond the maximum size that the License Manager can load. When this happens, the License Manager cannot start and the system will enter a continuous reboot loop. The Azurion system will not complete booting, and the message "X-ray system is booting..." remains displayed on the screen.Issue 2 - Longitudinal position error - Applicable only to azurion systems with Poly-G3 frontal stand. When requesting a movement of the frontal stand in longitudinal direction, the Azurion system may detect a mismatch between the expected longitudinal (set-) position and the measured (actual) position of the stand. This mismatch is caused by a glitch in the longitudinal position value provided by the position sensing potentiometer.When the mismatch is detected, the motion power to the frontal stand is automatically cut off. The message "Some stand movements are not available" is displayed to the user.In this situation, X-ray imaging and table movements will continue to function.This issue can be resolved with a cold restart of the system or by restarting the geometry.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Skus
Other
Model numbers
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Philips Medical Nederland B.v.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.