An official Health Canada notice concerns Philips azurion system. Philips has identified a safety issue affecting certain Azurion systems potentially leading to loss of imaging functionality. The power distribution unit is an internal assembly of the Philips Azurion system that contains multiple power supplies. Philips has identified that, in certain Azurion systems, one or more of these power supplies can fail due to energy peaks (surge voltages) during start or restart of the system.When a power supply failure occurs during a start or restart, the power distribution unit does not provide the required direct current power to essential system components. As a result, the system does not successfully complete the (re-)start, imaging functionality is unavailable, and the system remains unavailable for clinical use. Published 2026-10-29.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82876

Official notice

2026-10-29

The hazard

Philips has identified a safety issue affecting certain Azurion systems potentially leading to loss of imaging functionality. The power distribution unit is an internal assembly of the Philips Azurion system that contains multiple power supplies. Philips has identified that, in certain Azurion systems, one or more of these power supplies can fail due to energy peaks (surge voltages) during start or restart of the system.When a power supply failure occurs during a start or restart, the power distribution unit does not provide the required direct current power to essential system components. As a result, the system does not successfully complete the (re-)start, imaging functionality is unavailable, and the system remains unavailable for clinical use.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

722 224, 722 079, 722 067, 722 223, 722 078, 722 226, 722 068

Other

RA RA-82876

Model numbers

722 224, 722 079, 722 067, 722 223, 722 078, 722 226, 722 068

Recall Numbers

82876

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - Not applicable., Model or catalogue number - 722 224

Product 2

Lot or serial number - Not applicable., Model or catalogue number - 722 079

Product 3

Lot or serial number - Not applicable., Model or catalogue number - 722 067

Product 4

Lot or serial number - Not applicable., Model or catalogue number - 722 223

Product 5

Lot or serial number - Not applicable., Model or catalogue number - 722 078

Product 6

Lot or serial number - Not applicable., Model or catalogue number - 722 226

Product 7

Lot or serial number - Not applicable., Model or catalogue number - 722 068

Hazard

Philips has identified a safety issue affecting certain Azurion systems potentially leading to loss of imaging functionality. The power distribution unit is an internal assembly of the Philips Azurion system that contains multiple power supplies. Philips has identified that, in certain Azurion systems, one or more of these power supplies can fail due to energy peaks (surge voltages) during start or restart of the system.When a power supply failure occurs during a start or restart, the power distribution unit does not provide the required direct current power to essential system components. As a result, the system does not successfully complete the (re-)start, imaging functionality is unavailable, and the system remains unavailable for clinical use.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Philips Medical Nederland B.v.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82876

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 29 Oct 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button