An official Health Canada notice concerns Philips allura xper fd series. Philips has identified a potential safety issue affecting Allura R8.2 systems equipped with a AD7x patient table type potentially leading to loss of geometry movements. In some cases, following a system (re)start, manual and motorized geometry movements may be disabled due to a firmware issue in the Advanced Motion Controller (AMC) single-axis motion control unit used in the AD7x tables of Allura R8.2 systems. When the issue occurs, no message is displayed to the user. In this situation, X-ray imaging is available. The issue can be resolved by performing a cold restart of the system or by restarting the geometry. Published 2026-07-16.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82333

Official notice

2026-07-16

The hazard

Philips has identified a potential safety issue affecting Allura R8.2 systems equipped with a AD7x patient table type potentially leading to loss of geometry movements. In some cases, following a system (re)start, manual and motorized geometry movements may be disabled due to a firmware issue in the Advanced Motion Controller (AMC) single-axis motion control unit used in the AD7x tables of Allura R8.2 systems. When the issue occurs, no message is displayed to the user. In this situation, X-ray imaging is available. The issue can be resolved by performing a cold restart of the system or by restarting the geometry.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

722028, 722038, 722058, 722026

Other

RA RA-82333

Model numbers

722028, 722038, 722058, 722026

Recall Numbers

82333

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - Not applicable., Model or catalogue number - 722028

Product 2

Lot or serial number - Not applicable., Model or catalogue number - 722038

Product 3

Lot or serial number - Not applicable., Model or catalogue number - 722058

Product 4

Lot or serial number - Not applicable., Model or catalogue number - 722026

Hazard

Philips has identified a potential safety issue affecting Allura R8.2 systems equipped with a AD7x patient table type potentially leading to loss of geometry movements. In some cases, following a system (re)start, manual and motorized geometry movements may be disabled due to a firmware issue in the Advanced Motion Controller (AMC) single-axis motion control unit used in the AD7x tables of Allura R8.2 systems. When the issue occurs, no message is displayed to the user. In this situation, X-ray imaging is available. The issue can be resolved by performing a cold restart of the system or by restarting the geometry.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Philips Medical Nederland B.v.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82333

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 16 Jul 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button