Image not available in RecallScope.
Philips ingenia, achieva and intera mr systems
Official notice
9 Sept 2026
Last checked
13 Nov 2026
An official Health Canada notice concerns Philips ingenia, achieva and intera mr systems. Philips has become aware of a software issue affecting Philips Magnetic Resonance (MR) imaging systems operating with software versions R5 & R11, which may terminate an examination during the image reconstruction process. If the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory, the system will freeze, preventing image generation and requiring a system re-start. Published 2026-09-09.
What to do now
Contact the manufacturer if you require additionl information.
Recall number
82609
Official notice
2026-09-09
The hazard
The hazard
Philips has become aware of a software issue affecting Philips Magnetic Resonance (MR) imaging systems operating with software versions R5 & R11, which may terminate an examination during the image reconstruction process. If the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory, the system will freeze, preventing image generation and requiring a system re-start.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Skus
Other
Model numbers
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Product 2
Product 3
Product 4
Product 5
Product 6
Product 7
Product 8
Product 9
Product 10
Product 11
Product 12
Product 13
Product 14
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Philips Medical Nederland B.v.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.