Open official notice

Philips ingenia, achieva and intera mr systems

Official notice

9 Sept 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Philips ingenia, achieva and intera mr systems. Philips has become aware of a software issue affecting Philips Magnetic Resonance (MR) imaging systems operating with software versions R5 & R11, which may terminate an examination during the image reconstruction process. If the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory, the system will freeze, preventing image generation and requiring a system re-start. Published 2026-09-09.

What to do now

Contact the manufacturer if you require additionl information.

Recall number

82609

Official notice

2026-09-09

The hazard

Philips has become aware of a software issue affecting Philips Magnetic Resonance (MR) imaging systems operating with software versions R5 & R11, which may terminate an examination during the image reconstruction process. If the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory, the system will freeze, preventing image generation and requiring a system re-start.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

781315, 781271, 782108, 782139, 782109, 781356, 782138, 782107, 781358

Other

RA RA-82609

Model numbers

781315, 781271, 782108, 782139, 782109, 781356, 782138, 782107, 781358

Recall Numbers

82609

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - Not applicable., Model or catalogue number - 781315

Product 2

Lot or serial number - Not applicable., Model or catalogue number - 1.5T

Product 3

Lot or serial number - Not applicable., Model or catalogue number - 781271

Product 4

Lot or serial number - Not applicable., Model or catalogue number - ACHIEVA 1.5T

Product 5

Lot or serial number - Not applicable., Model or catalogue number - ACHIEVA 3.0T

Product 6

Lot or serial number - Not applicable., Model or catalogue number - 782108

Product 7

Lot or serial number - Not applicable., Model or catalogue number - 782139

Product 8

Lot or serial number - Not applicable., Model or catalogue number - 3.0T

Product 9

Lot or serial number - Not applicable., Model or catalogue number - 782109

Product 10

Lot or serial number - Not applicable., Model or catalogue number - 781356

Product 11

Lot or serial number - Not applicable., Model or catalogue number - 782138

Product 12

Lot or serial number - Not applicable., Model or catalogue number - 782107

Product 13

Lot or serial number - Not applicable., Model or catalogue number - 781358

Product 14

Lot or serial number - Not applicable., Model or catalogue number - INTERA 1.5T

Hazard

Philips has become aware of a software issue affecting Philips Magnetic Resonance (MR) imaging systems operating with software versions R5 & R11, which may terminate an examination during the image reconstruction process. If the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory, the system will freeze, preventing image generation and requiring a system re-start.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Philips Medical Nederland B.v.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82609

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 9 Sept 2026. The next one won’t wait for you to search.