Astral ventilators

Official notice

13 Oct 2023

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Astral ventilators. Resmed's Astral ventilator is designed to deliver a Total Power Failure (TPF) audible alarm for at least 2 minutes to alert the carer/clinician when the ventilator experiences a total loss of power that results in a stop of ventilation. The TPF audible alarm is powered by an electrical component known as a supercapacitor. This component degrades over time, which may cause the TPF alarm to sound for less than 2 minutes, or in some cases, to not sound at all. Recall start date: September 13, 2023 Published 2023-10-13.

What to do now

Manufacturer provided instructions on how to response. Will implement corrections when available.

Recall number

74413

Official notice

2023-10-13

The hazard

Resmed's Astral ventilator is designed to deliver a Total Power Failure (TPF) audible alarm for at least 2 minutes to alert the carer/clinician when the ventilator experiences a total loss of power that results in a stop of ventilation. The TPF audible alarm is powered by an electrical component known as a supercapacitor. This component degrades over time, which may cause the TPF alarm to sound for less than 2 minutes, or in some cases, to not sound at all. Recall start date: September 13, 2023

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

74413

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - Astral Ventilators, Lot or serial number - Contact the manufacturer., Model or catalog number - 27003

Product 2

Affected products - Astral Ventilators, Lot or serial number - Contact the manufacturer., Model or catalog number - 27001

Hazard

Resmed's Astral ventilator is designed to deliver a Total Power Failure (TPF) audible alarm for at least 2 minutes to alert the carer/clinician when the ventilator experiences a total loss of power that results in a stop of ventilation. The TPF audible alarm is powered by an electrical component known as a supercapacitor. This component degrades over time, which may cause the TPF alarm to sound for less than 2 minutes, or in some cases, to not sound at all. Recall start date: September 13, 2023

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Resmed, Astral Ventilators.

Source & status

Tier

official

Category

Health

Official classification

Type II

Recall details

Recall number

74413

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 13 Oct 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button