Embletta mpr pg

Official notice

6 Oct 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Embletta mpr pg. Natus has become aware of complaints related to certain Embletta MPR devices where the device is unable to correctly format its internal memory. When the internal memory cannot be correctly formatted, previously recorded studies may remain on the device and may be updated with incorrect patient information from a subsequent study. The issue may also prevent studies from initiating recording successfully or may cause recordings to end prematurely, resulting in incomplete or unavailable diagnostic data and the potential need to repeat a patient study. In some cases, users may be unable to download previously recorded studies from the device. Customers are unable to correct the memory formatting issue through normal device operation, and there is currently no workaround available. These conditions may increase the risk of incorrect patient information being associated with a study, loss of diagnostic data, delays in diagnosis, or the need to repeat patient testing. Published 2026-10-06.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82736

Official notice

2026-10-06

The hazard

Natus has become aware of complaints related to certain Embletta MPR devices where the device is unable to correctly format its internal memory. When the internal memory cannot be correctly formatted, previously recorded studies may remain on the device and may be updated with incorrect patient information from a subsequent study. The issue may also prevent studies from initiating recording successfully or may cause recordings to end prematurely, resulting in incomplete or unavailable diagnostic data and the potential need to repeat a patient study. In some cases, users may be unable to download previously recorded studies from the device. Customers are unable to correct the memory formatting issue through normal device operation, and there is currently no workaround available. These conditions may increase the risk of incorrect patient information being associated with a study, loss of diagnostic data, delays in diagnosis, or the need to repeat patient testing.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

EBMPG-C1013446, EBMPG-C1013450, EBMPG-C1013674, EBMPG-C1013449, EBMPG-C1013457, EBMPG-C1013677, EBMPG-C1013435, EBMPG-C1013451

Skus

2003011

Other

RA RA-82736

Model numbers

2003011

Recall Numbers

82736

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - EBMPG-C1013446, Model or catalogue number - 2003011

Product 2

Lot or serial number - EBMPG-C1013450, Model or catalogue number - 2003011

Product 3

Lot or serial number - EBMPG-C1013674, Model or catalogue number - 2003011

Product 4

Lot or serial number - EBMPG-C1013449, Model or catalogue number - 2003011

Product 5

Lot or serial number - EBMPG-C1013457, Model or catalogue number - 2003011

Product 6

Lot or serial number - EBMPG-C1013677, Model or catalogue number - 2003011

Product 7

Lot or serial number - EBMPG-C1013435, Model or catalogue number - 2003011

Product 8

Lot or serial number - EBMPG-C1013451, Model or catalogue number - 2003011

Hazard

Natus has become aware of complaints related to certain Embletta MPR devices where the device is unable to correctly format its internal memory. When the internal memory cannot be correctly formatted, previously recorded studies may remain on the device and may be updated with incorrect patient information from a subsequent study. The issue may also prevent studies from initiating recording successfully or may cause recordings to end prematurely, resulting in incomplete or unavailable diagnostic data and the potential need to repeat a patient study. In some cases, users may be unable to download previously recorded studies from the device. Customers are unable to correct the memory formatting issue through normal device operation, and there is currently no workaround available. These conditions may increase the risk of incorrect patient information being associated with a study, loss of diagnostic data, delays in diagnosis, or the need to repeat patient testing.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada · Ontario

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Natus Medical Adba Excel-tech.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82736

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 6 Oct 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button