Open official notice

Blanketrol® ii 115v and blanketrol® iii 115v

Official notice

28 Aug 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Blanketrol® ii 115v and blanketrol® iii 115v. Gentherm Medical has received complaints of overheated circuit boards in 0.53% of distributed devices. Overheating can cause the circuit board to stop operating, which may result in a device that shuts down at startup or during use on a patient. Based on the low probability of harm, and absence of reported harm, associated with this issue, users may continue to use the device. In an abundance of caution, it is recommended to have an alternative means of providing therapy available. Published 2026-08-28.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82413

Official notice

2026-08-28

The hazard

Gentherm Medical has received complaints of overheated circuit boards in 0.53% of distributed devices. Overheating can cause the circuit board to stop operating, which may result in a device that shuts down at startup or during use on a patient. Based on the low probability of harm, and absence of reported harm, associated with this issue, users may continue to use the device. In an abundance of caution, it is recommended to have an alternative means of providing therapy available.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

151-2-01474, 151-2-01472, 114-2-00258, 131-2-00535, 124-2-00447, 124-2-00490, 114-2-00259

Skus

222S, 233

Other

RA RA-82413

Model numbers

222S, 233

Recall Numbers

82413

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - 151-2-01474, Model or catalogue number - 222S

Product 2

Lot or serial number - 151-2-01472, Model or catalogue number - 222S

Product 3

Lot or serial number - 114-2-00258, Model or catalogue number - 222S

Product 4

Lot or serial number - 131-2-00535, Model or catalogue number - 222S

Product 5

Lot or serial number - 124-2-00447, Model or catalogue number - 222S

Product 6

Lot or serial number - 124-2-00490, Model or catalogue number - 222S

Product 7

Lot or serial number - 114-2-00259, Model or catalogue number - 222S

Product 8

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 233

Hazard

Gentherm Medical has received complaints of overheated circuit boards in 0.53% of distributed devices. Overheating can cause the circuit board to stop operating, which may result in a device that shuts down at startup or during use on a patient. Based on the low probability of harm, and absence of reported harm, associated with this issue, users may continue to use the device. In an abundance of caution, it is recommended to have an alternative means of providing therapy available.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Gentherm Medical.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82413

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 28 Aug 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button