Automated impella controller (aic)

Official notice

27 Aug 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Automated impella controller (aic). Abiomed has initiated a voluntary medical device recall (removal) after identification of an increase in complaints related to purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly. The cause of the purge flag failures is a combination of dextrose buildup from a purge cassette leak and material embrittlement from usage, aging, or exposure to cleaning chemicals. This issue may occur from usage and is not a manufacturing defect. The purge flag component is present in all manufactured Automated Impella Controller (AIC) in Canada. However, product may continue to be used to ensure continuity of care and a prioritized risk-based strategy for component replacement will be implemented. The purge flag is the component that engages with the purge disc and allows the AIC to recognize the purge disc. These AIC purge flag component failures may result most frequently in user inconvenience due to troubleshooting, attempted purge cassette exchange and AIC exchange. If the failure occurs before initiation of support, exposure may result in a delay to procedure, which is expected to remain clinically manageable in most cases. In the case of an urgent need for mechanical circulatory support or if the failure occurs during active Impella support, exposure may result in a temporary period of inadequate hemodynamic support during controller transition. Published 2026-08-27.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82534

Official notice

2026-08-27

The hazard

Abiomed has initiated a voluntary medical device recall (removal) after identification of an increase in complaints related to purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly. The cause of the purge flag failures is a combination of dextrose buildup from a purge cassette leak and material embrittlement from usage, aging, or exposure to cleaning chemicals. This issue may occur from usage and is not a manufacturing defect. The purge flag component is present in all manufactured Automated Impella Controller (AIC) in Canada. However, product may continue to be used to ensure continuity of care and a prioritized risk-based strategy for component replacement will be implemented. The purge flag is the component that engages with the purge disc and allows the AIC to recognize the purge disc. These AIC purge flag component failures may result most frequently in user inconvenience due to troubleshooting, attempted purge cassette exchange and AIC exchange. If the failure occurs before initiation of support, exposure may result in a delay to procedure, which is expected to remain clinically manageable in most cases. In the case of an urgent need for mechanical circulatory support or if the failure occurs during active Impella support, exposure may result in a temporary period of inadequate hemodynamic support during controller transition.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

0042-0040-CA, 0042-0000-US, 0042-0000-CA

Other

RA RA-82534

Model numbers

0042-0040-CA, 0042-0000-US, 0042-0000-CA

Recall Numbers

82534

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - All lots., Model or catalogue number - 0042-0040-CA

Product 2

Lot or serial number - All lots., Model or catalogue number - 0042-0000-US

Product 3

Lot or serial number - All lots., Model or catalogue number - 0042-0000-CA

Hazard

Abiomed has initiated a voluntary medical device recall (removal) after identification of an increase in complaints related to purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly. The cause of the purge flag failures is a combination of dextrose buildup from a purge cassette leak and material embrittlement from usage, aging, or exposure to cleaning chemicals. This issue may occur from usage and is not a manufacturing defect. The purge flag component is present in all manufactured Automated Impella Controller (AIC) in Canada. However, product may continue to be used to ensure continuity of care and a prioritized risk-based strategy for component replacement will be implemented. The purge flag is the component that engages with the purge disc and allows the AIC to recognize the purge disc. These AIC purge flag component failures may result most frequently in user inconvenience due to troubleshooting, attempted purge cassette exchange and AIC exchange. If the failure occurs before initiation of support, exposure may result in a delay to procedure, which is expected to remain clinically manageable in most cases. In the case of an urgent need for mechanical circulatory support or if the failure occurs during active Impella support, exposure may result in a temporary period of inadequate hemodynamic support during controller transition.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Manufacturer

Not provided

Phone

Not provided

Brand: Abiomed.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82534

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 27 Aug 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button