Automated impella controller

Official notice

25 Feb 2026

Last checked

25 Feb 2026

High riskHealth Canada
Watch Abiomed

An official Health Canada notice concerns Automated impella controller. Abiomed has identified that Automated Impella Controller (AICS) with software V8.2.3, when used in conjunction with an Impella 5.5 with smart assist set (product code 0550-0004), has a delayed "purge system blocked" alarm. Software versions post-v10.1 have this issue addressed. Abiomed recommends that you do not use the AIC with an Impella 5.5 with smart assist set (product code 0550-0004) unless AIC software is updated to v10.1 or above. Published 2026-02-25.

What to do now

Contact the manufacturer if you require additional information.

Recall number

81652

Official notice

2026-02-25

The hazard

Abiomed has identified that Automated Impella Controller (AICS) with software V8.2.3, when used in conjunction with an Impella 5.5 with smart assist set (product code 0550-0004), has a delayed "purge system blocked" alarm. Software versions post-v10.1 have this issue addressed. Abiomed recommends that you do not use the AIC with an Impella 5.5 with smart assist set (product code 0550-0004) unless AIC software is updated to v10.1 or above.

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Skus

0042-0040-CA, 0042-0000-CA

Other

RA RA-81652

Model Numbers

0042-0040-CA, 0042-0000-CA

Recall Numbers

81652

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Lot or serial number - Software version 8.2.3, Model or catalogue number - 0042-0040-CA

Product 2

Lot or serial number - Software version 8.2.3, Model or catalogue number - 0042-0000-CA

Risk / hazard

Abiomed has identified that Automated Impella Controller (AICS) with software V8.2.3, when used in conjunction with an Impella 5.5 with smart assist set (product code 0550-0004), has a delayed "purge system blocked" alarm. Software versions post-v10.1 have this issue addressed. Abiomed recommends that you do not use the AIC with an Impella 5.5 with smart assist set (product code 0550-0004) unless AIC software is updated to v10.1 or above.

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Abiomed

Manufacturer

Not provided

Phone

Not provided

Brand: Abiomed.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type I

App estimate

high

Recall details

Recall number

81652

Classification

Type I

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Abiomed introducer kits with impella heart pumps

High riskHealth Canada29 Jun 2026
CA
Recall image placeholder

Impella controller

Critical riskHealth Canada13 May 2026
CA
Recall image placeholder

Automated impella® controller

Critical riskHealth Canada28 Apr 2026
CA
Recall image placeholder

Abiomed purge cassette generation 1 for impella

High riskHealth Canada4 Mar 2026
CA
Recall image placeholder

Automated impella® controller

Medium riskHealth Canada23 Oct 2025
CA
Recall image placeholder

Impella controller

Critical riskHealth Canada18 Sept 2025
CA
Recall image placeholder

Impella® controller

Critical riskHealth Canada29 Aug 2025
CA
Recall image placeholder

Impella® controller

High riskHealth Canada9 Jul 2025
CA
Recall image placeholder

Impella cp with smartassist and impella 5.5 with smartassist

High riskHealth Canada5 Sept 2024
CA
Recall image placeholder

Impella cp

High riskHealth Canada20 Mar 2024
CA
Recall image placeholder

Impella products

Medium riskHealth Canada12 Mar 2024
CA
Recall image placeholder

Impella systems

High riskHealth Canada6 Mar 2024

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 25 Feb 2026. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.