No image published for this recall

Automated impella controller

Official notice

25 Feb 2026

HighHealth Canada

An official Health Canada notice concerns Automated impella controller. Abiomed has identified that Automated Impella Controller (AICS) with software V8.2.3, when used in conjunction with an Impella 5.5 with smart assist set (product code 0550-0004), has a delayed "purge system blocked" alarm. Software versions post-v10.1 have this issue addressed. Abiomed recommends that you do not use the AIC with an Impella 5.5 with smart assist set (product code 0550-0004) unless AIC software is updated to v10.1 or above. Published 2026-02-25.

What to do now

Contact the manufacturer if you require additional information.

Recall number

81652

Official notice

2026-02-25

The hazard

Abiomed has identified that Automated Impella Controller (AICS) with software V8.2.3, when used in conjunction with an Impella 5.5 with smart assist set (product code 0550-0004), has a delayed "purge system blocked" alarm. Software versions post-v10.1 have this issue addressed. Abiomed recommends that you do not use the AIC with an Impella 5.5 with smart assist set (product code 0550-0004) unless AIC software is updated to v10.1 or above.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Skus

0042-0040-CA, 0042-0000-CA

Other

RA RA-81652

Model Numbers

0042-0040-CA, 0042-0000-CA

Recall Numbers

81652

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - Software version 8.2.3, Model or catalogue number - 0042-0040-CA

Product 2

Lot or serial number - Software version 8.2.3, Model or catalogue number - 0042-0000-CA

Risk / hazard

Abiomed has identified that Automated Impella Controller (AICS) with software V8.2.3, when used in conjunction with an Impella 5.5 with smart assist set (product code 0550-0004), has a delayed "purge system blocked" alarm. Software versions post-v10.1 have this issue addressed. Abiomed recommends that you do not use the AIC with an Impella 5.5 with smart assist set (product code 0550-0004) unless AIC software is updated to v10.1 or above.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Manufacturer

Not provided

Phone

Not provided

Brand: Abiomed.

Source & status

Source

Health Canada

Tier

beta

Category

HEALTH

Classification

Type I

App estimate

high

Recall details

Recall number

81652

Classification

Type I

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

HEALTH
High

Abiomed introducer kits with impella heart pumps

Health Canada29 Jun 2026

Brand match

High risk

Markets

CA
HEALTH
Critical

Impella controller

Health Canada13 May 2026

Brand match

Critical risk

Markets

CA
HEALTH
Critical

Automated impella® controller

Health Canada28 Apr 2026

Brand match

Critical risk

Markets

CA
HEALTH
High

Abiomed purge cassette generation 1 for impella

Health Canada4 Mar 2026

Brand match

High risk

Markets

CA
HEALTH
Medium

Automated impella® controller

Health Canada23 Oct 2025

Brand match

Medium risk

Markets

CA
HEALTH
Critical

Impella controller

Health Canada18 Sept 2025

Brand match

Critical risk

Markets

CA
HEALTH
Critical

Impella® controller

Health Canada29 Aug 2025

Brand match

Critical risk

Markets

CA
HEALTH
High

Impella® controller

Health Canada9 Jul 2025

Brand match

High risk

Markets

CA
HEALTH
High

Impella cp with smartassist and impella 5.5 with smartassist

Health Canada5 Sept 2024

Brand match

High risk

Markets

CA
HEALTH
High

Impella cp

Health Canada20 Mar 2024

Brand match

High risk

Markets

CA
HEALTH
Medium

Impella products

Health Canada12 Mar 2024

Brand match

Medium risk

Markets

CA
HEALTH
High

Impella systems

Health Canada6 Mar 2024

Brand match

High risk

Markets

CA

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 25 Feb 2026. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.