Automated impella® controllers (aic)

Official notice

2 Oct 2026

Last checked

13 Nov 2026

RecallScope risk estimate:HighHealth Canada

An official Health Canada notice concerns Automated impella® controllers (aic). Abiomed has decided to initiate a voluntary medical device recall (removal) for design changes that were previously implemented through the Abiomed field service process. Abiomed performed a retrospective review of Field Change Notices (FCN) and determined that certain AIC consoles in the field did not receive the updates introduced by FCN 113.Failure to implement the applicable FCN 113 updates may result in the screen freezing during the Purge Cassette Change Wizard and the display of a false controller error.These defects could delay the initiation of support during device set-up and, in high-risk patient populations, may lead to temporary inadequate hemodynamic support or a temporary loss of hemodynamic support. Hospital inventory may continue to be used to ensure continuity of care. Not all units are affected by this issue. Published 2026-10-02.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82721

Official notice

2026-10-02

The hazard

Abiomed has decided to initiate a voluntary medical device recall (removal) for design changes that were previously implemented through the Abiomed field service process. Abiomed performed a retrospective review of Field Change Notices (FCN) and determined that certain AIC consoles in the field did not receive the updates introduced by FCN 113.Failure to implement the applicable FCN 113 updates may result in the screen freezing during the Purge Cassette Change Wizard and the display of a false controller error.These defects could delay the initiation of support during device set-up and, in high-risk patient populations, may lead to temporary inadequate hemodynamic support or a temporary loss of hemodynamic support. Hospital inventory may continue to be used to ensure continuity of care. Not all units are affected by this issue.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

IC13503, IC13266, IC11171, IC11172, IC1769, IC5821, IC5824, IC9416, IC9414, IC1179

Skus

0042-0040-CA, 0042-0000-CA, 0042-0000-US

Other

RA RA-82721

Model numbers

0042-0040-CA, 0042-0000-CA, 0042-0000-US

Recall Numbers

82721

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - IC13503, Model or catalogue number - 0042-0040-CA

Product 2

Lot or serial number - IC13266, Model or catalogue number - 0042-0040-CA

Product 3

Lot or serial number - IC11171, Model or catalogue number - 0042-0040-CA

Product 4

Lot or serial number - IC11172, Model or catalogue number - 0042-0040-CA

Product 5

Lot or serial number - IC1769, Model or catalogue number - 0042-0000-CA

Product 6

Lot or serial number - IC5821, Model or catalogue number - 0042-0000-CA

Product 7

Lot or serial number - IC5824, Model or catalogue number - 0042-0000-CA

Product 8

Lot or serial number - IC9416, Model or catalogue number - 0042-0000-CA

Product 9

Lot or serial number - IC9414, Model or catalogue number - 0042-0000-CA

Product 10

Lot or serial number - IC1179, Model or catalogue number - 0042-0000-US

Hazard

Abiomed has decided to initiate a voluntary medical device recall (removal) for design changes that were previously implemented through the Abiomed field service process. Abiomed performed a retrospective review of Field Change Notices (FCN) and determined that certain AIC consoles in the field did not receive the updates introduced by FCN 113.Failure to implement the applicable FCN 113 updates may result in the screen freezing during the Purge Cassette Change Wizard and the display of a false controller error.These defects could delay the initiation of support during device set-up and, in high-risk patient populations, may lead to temporary inadequate hemodynamic support or a temporary loss of hemodynamic support. Hospital inventory may continue to be used to ensure continuity of care. Not all units are affected by this issue.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Manufacturer

Not provided

Phone

Not provided

Brand: Abiomed.

Source & status

Tier

beta

Category

Health

Official classification

Type I

Recall details

Recall number

82721

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 2 Oct 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button