No image published for this recall
Philips incisive ct
Official notice
11 May 2023
Last checked
11 May 2023
An official Health Canada notice concerns Philips incisive ct. Philips has identified a hardware issue with a metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the incisive CT system. This component may become detached and make contact with other components located within the incisive CT during rotation. Other components could be damaged due to the contact with the detached component. If this issue occurs, a loud noise will be emitted and the system will shut down. The cover may become displaced if the following sequence occurs: • Damaged component(s) come loose • The damaged component(s) contact other components in the system and cause them to break • A broken component makes contact with the top right gantry cover The displaced cover could create a small gap allowing a fragment of a damaged component to be expelled at a low velocity. The maximum weight of a fragment that could be expelled is 60g (2.12 ounces). Recall start date: May 1, 2023 Published 2023-05-11.
What to do now
Contact the manufacturer if you require additional information.
Recall number
73644
Official notice
2023-05-11
The hazard
The hazard
Philips has identified a hardware issue with a metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the incisive CT system. This component may become detached and make contact with other components located within the incisive CT during rotation. Other components could be damaged due to the contact with the detached component. If this issue occurs, a loud noise will be emitted and the system will shut down. The cover may become displaced if the following sequence occurs: • Damaged component(s) come loose • The damaged component(s) contact other components in the system and cause them to break • A broken component makes contact with the top right gantry cover The displaced cover could create a small gap allowing a fragment of a damaged component to be expelled at a low velocity. The maximum weight of a fragment that could be expelled is 60g (2.12 ounces). Recall start date: May 1, 2023
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Philips.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Hard disk drives in azurion and allura systems

Azurion systems

Azurion and allura xper systems

Ingenia and achieva systems

Aneurysmflow

Philips allura xper systems

Trilogy evo and ev300

Allura xper and allura cv20 systems

Patient information center ix

Azurion r2.1.10 and r2.2.10 systems

Achieva 3.0t mri system-main unit

Philips lumify systems
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 11 May 2023. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.