Philips incisive ct

Official notice

11 May 2023

Last checked

11 May 2023

Medium riskHealth Canada

An official Health Canada notice concerns Philips incisive ct. Philips has identified a hardware issue with a metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the incisive CT system. This component may become detached and make contact with other components located within the incisive CT during rotation. Other components could be damaged due to the contact with the detached component. If this issue occurs, a loud noise will be emitted and the system will shut down. The cover may become displaced if the following sequence occurs: • Damaged component(s) come loose • The damaged component(s) contact other components in the system and cause them to break • A broken component makes contact with the top right gantry cover The displaced cover could create a small gap allowing a fragment of a damaged component to be expelled at a low velocity. The maximum weight of a fragment that could be expelled is 60g (2.12 ounces). Recall start date: May 1, 2023 Published 2023-05-11.

What to do now

Contact the manufacturer if you require additional information.

Recall number

73644

Official notice

2023-05-11

The hazard

Philips has identified a hardware issue with a metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the incisive CT system. This component may become detached and make contact with other components located within the incisive CT during rotation. Other components could be damaged due to the contact with the detached component. If this issue occurs, a loud noise will be emitted and the system will shut down. The cover may become displaced if the following sequence occurs: • Damaged component(s) come loose • The damaged component(s) contact other components in the system and cause them to break • A broken component makes contact with the top right gantry cover The displaced cover could create a small gap allowing a fragment of a damaged component to be expelled at a low velocity. The maximum weight of a fragment that could be expelled is 60g (2.12 ounces). Recall start date: May 1, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

73644

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Incisive Ct, Lot or serial number - All lots., Model or catalog number - 728143

Risk / hazard

Philips has identified a hardware issue with a metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the incisive CT system. This component may become detached and make contact with other components located within the incisive CT during rotation. Other components could be damaged due to the contact with the detached component. If this issue occurs, a loud noise will be emitted and the system will shut down. The cover may become displaced if the following sequence occurs: • Damaged component(s) come loose • The damaged component(s) contact other components in the system and cause them to break • A broken component makes contact with the top right gantry cover The displaced cover could create a small gap allowing a fragment of a damaged component to be expelled at a low velocity. The maximum weight of a fragment that could be expelled is 60g (2.12 ounces). Recall start date: May 1, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Manufacturer

Not provided

Phone

Not provided

Brand: Philips.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

73644

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
No image published for this recall

Hard disk drives in azurion and allura systems

Critical riskHealth Canada15 May 2026
CA
No image published for this recall

Azurion systems

Critical riskHealth Canada7 May 2026
CA
No image published for this recall

Azurion and allura xper systems

Medium riskHealth Canada6 May 2026
CA
No image published for this recall

Ingenia and achieva systems

Critical riskHealth Canada16 Apr 2026
CA
No image published for this recall

Aneurysmflow

Medium riskHealth Canada2 Apr 2026
CA
No image published for this recall

Philips allura xper systems

Medium riskHealth Canada4 Jun 2026
CA
No image published for this recall

Trilogy evo and ev300

High riskHealth Canada18 Feb 2026
CA
No image published for this recall

Allura xper and allura cv20 systems

Critical riskHealth Canada6 Jan 2026
CA
No image published for this recall

Patient information center ix

Medium riskHealth Canada31 Dec 2025
CA
No image published for this recall

Azurion r2.1.10 and r2.2.10 systems

Critical riskHealth Canada29 Dec 2025
CA
No image published for this recall

Achieva 3.0t mri system-main unit

Medium riskHealth Canada18 Dec 2025
CA
No image published for this recall

Philips lumify systems

Medium riskHealth Canada26 Nov 2025

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 11 May 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.