Azurion and allura xper systems

Official notice

6 May 2026

Last checked

6 May 2026

Medium riskHealth Canada
Watch Philips

An official Health Canada notice concerns Azurion and allura xper systems. The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the x-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message "low load fluoroscopy flavor selected: tube cooler problem". This condition cannot be resolved by restarting the system. Under normal conditions, any leaked oil is contained within the system. In rare cases of more significant leakage, oil may extend beyond the intended containment area. If oil then comes into contact with hot surfaces, fumes may be generated. These fumes may activate environmental detection systems. Published 2026-05-06.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82040

Official notice

2026-05-06

The hazard

The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the x-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message "low load fluoroscopy flavor selected: tube cooler problem". This condition cannot be resolved by restarting the system. Under normal conditions, any leaked oil is contained within the system. In rare cases of more significant leakage, oil may extend beyond the intended containment area. If oil then comes into contact with hot surfaces, fumes may be generated. These fumes may activate environmental detection systems.

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Skus

722 068, ALLURA XPER FD10, 001443, 722028, 722 078, 722 079, 722026, 722029, 722035, 722 067

Other

RA RA-82040

Model Numbers

722 068, ALLURA XPER FD10, 001443, 722028, 722 078, 722 079, 722026, 722029, 722035, 722 067

Recall Numbers

82040

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Lot or serial number - Not applicable., Model or catalogue number - 722 068

Product 2

Lot or serial number - Not applicable., Model or catalogue number - ALLURA XPER FD10

Product 3

Lot or serial number - Not applicable., Model or catalogue number - 001443

Product 4

Lot or serial number - Not applicable., Model or catalogue number - 722028

Product 5

Lot or serial number - Not applicable., Model or catalogue number - 722 078

Product 6

Lot or serial number - Not applicable., Model or catalogue number - 722 079

Product 7

Lot or serial number - Not applicable., Model or catalogue number - 722026

Product 8

Lot or serial number - Not applicable., Model or catalogue number - 722029

Product 9

Lot or serial number - Not applicable., Model or catalogue number - 722035

Product 10

Lot or serial number - Not applicable., Model or catalogue number - 722 067

Risk / hazard

The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the x-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message "low load fluoroscopy flavor selected: tube cooler problem". This condition cannot be resolved by restarting the system. Under normal conditions, any leaked oil is contained within the system. In rare cases of more significant leakage, oil may extend beyond the intended containment area. If oil then comes into contact with hot surfaces, fumes may be generated. These fumes may activate environmental detection systems.

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Philips

Manufacturer

Not provided

Phone

Not provided

Brand: Philips.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

82040

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Philips allura xper systems

Medium riskHealth Canada4 Jun 2026
CA
Recall image placeholder

Hard disk drives in azurion and allura systems

Critical riskHealth Canada15 May 2026
CA
Recall image placeholder

Azurion systems

Critical riskHealth Canada7 May 2026
CA
Recall image placeholder

Ingenia and achieva systems

Critical riskHealth Canada16 Apr 2026
CA
Recall image placeholder

Aneurysmflow

Medium riskHealth Canada2 Apr 2026
CA
Recall image placeholder

Allura xper and azurion systems

Medium riskHealth Canada17 Mar 2026
CA
Recall image placeholder

Intellivue mx patient monitor system-intellibridge module ec10

Medium riskHealth Canada11 Mar 2026
CA
Recall image placeholder

Trilogy evo and ev300

High riskHealth Canada18 Feb 2026
CA
Recall image placeholder

Allura xper and allura cv20 systems

Critical riskHealth Canada6 Jan 2026
CA
Recall image placeholder

Azurion r2.1.10 and r2.2.10 systems

Critical riskHealth Canada29 Dec 2025
CA
Recall image placeholder

Achieva 3.0t mri system-main unit

Medium riskHealth Canada18 Dec 2025
CA
Recall image placeholder

Allura xper systems

Medium riskHealth Canada24 Sept 2025

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 6 May 2026. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.