An official Health Canada notice concerns Azurion and allura xper systems. The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the x-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message "low load fluoroscopy flavor selected: tube cooler problem". This condition cannot be resolved by restarting the system. Under normal conditions, any leaked oil is contained within the system. In rare cases of more significant leakage, oil may extend beyond the intended containment area. If oil then comes into contact with hot surfaces, fumes may be generated. These fumes may activate environmental detection systems. Published 2026-05-06.
What to do now
Contact the manufacturer if you require additional information.
Recall number
82040
Official notice
2026-05-06
The hazard
The hazard
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the x-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message "low load fluoroscopy flavor selected: tube cooler problem". This condition cannot be resolved by restarting the system. Under normal conditions, any leaked oil is contained within the system. In rare cases of more significant leakage, oil may extend beyond the intended containment area. If oil then comes into contact with hot surfaces, fumes may be generated. These fumes may activate environmental detection systems.
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Skus
Other
Model Numbers
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Product 2
Product 3
Product 4
Product 5
Product 6
Product 7
Product 8
Product 9
Product 10
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Philips.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Philips allura xper systems
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Hard disk drives in azurion and allura systems
Brand match
Critical risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Azurion systems
Brand match
Critical risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Ingenia and achieva systems
Brand match
Critical risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Aneurysmflow
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Trilogy evo and ev300
Brand match
High risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Allura xper and allura cv20 systems
Brand match
Critical risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Azurion r2.1.10 and r2.2.10 systems
Brand match
Critical risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Achieva 3.0t mri system-main unit
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Allura xper systems
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Pinnacle³ radiation therapy planning system
Brand match
Critical risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Allura and azurion systems
Brand match
Critical risk
Markets
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 6 May 2026. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.