Watch Philips

An official Health Canada notice concerns Azurion systems. Philips has identified that the float tabletop control on the control module of Azurion systems may experience mechanical wear over time. The float tabletop control is located on the control module at the table-side. When the control is actuated, the table brakes disengage, allowing adjustment of the longitudinal or transverse position of the table. Over time, mechanical wear may lead to detachment, cracking, or breakage of the float tabletop control, which may lead to reduced availability or unavailability of longitudinal and transverse tabletop movement, as well as the formation of sharp edges. Published 2026-05-07.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82045

Official notice

2026-05-07

The hazard

Philips has identified that the float tabletop control on the control module of Azurion systems may experience mechanical wear over time. The float tabletop control is located on the control module at the table-side. When the control is actuated, the table brakes disengage, allowing adjustment of the longitudinal or transverse position of the table. Over time, mechanical wear may lead to detachment, cracking, or breakage of the float tabletop control, which may lead to reduced availability or unavailability of longitudinal and transverse tabletop movement, as well as the formation of sharp edges.

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Skus

722 078, 722 067, 722 224, 722 079

Other

RA RA-82045

Model Numbers

722 078, 722 067, 722 224, 722 079

Recall Numbers

82045

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Lot or serial number - Not applicable., Model or catalogue number - 722 078

Product 2

Lot or serial number - Not applicable., Model or catalogue number - 722 067

Product 3

Lot or serial number - Not applicable., Model or catalogue number - 722 224

Product 4

Lot or serial number - Not applicable., Model or catalogue number - 722 079

Risk / hazard

Philips has identified that the float tabletop control on the control module of Azurion systems may experience mechanical wear over time. The float tabletop control is located on the control module at the table-side. When the control is actuated, the table brakes disengage, allowing adjustment of the longitudinal or transverse position of the table. Over time, mechanical wear may lead to detachment, cracking, or breakage of the float tabletop control, which may lead to reduced availability or unavailability of longitudinal and transverse tabletop movement, as well as the formation of sharp edges.

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Philips

Manufacturer

Not provided

Phone

Not provided

Brand: Philips.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

critical

Recall details

Recall number

82045

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Hard disk drives in azurion and allura systems

Critical riskHealth Canada15 May 2026
CA
Recall image placeholder

Azurion and allura xper systems

Medium riskHealth Canada6 May 2026
CA
Recall image placeholder

Ingenia and achieva systems

Critical riskHealth Canada16 Apr 2026
CA
Recall image placeholder

Aneurysmflow

Medium riskHealth Canada2 Apr 2026
CA
Recall image placeholder

Allura xper and azurion systems

Medium riskHealth Canada17 Mar 2026
CA
Recall image placeholder

Intellivue mx patient monitor system-intellibridge module ec10

Medium riskHealth Canada11 Mar 2026
CA
Recall image placeholder

Philips allura xper systems

Medium riskHealth Canada4 Jun 2026
CA
Recall image placeholder

Allura xper and allura cv20 systems

Critical riskHealth Canada6 Jan 2026
CA
Recall image placeholder

Patient information center ix

Medium riskHealth Canada31 Dec 2025
CA
Recall image placeholder

Azurion r2.1.10 and r2.2.10 systems

Critical riskHealth Canada29 Dec 2025
CA
Recall image placeholder

Achieva 3.0t mri system-main unit

Medium riskHealth Canada18 Dec 2025
CA
Recall image placeholder

Azurion r3.0 systems

Medium riskHealth Canada18 Dec 2025

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 7 May 2026. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.