Pds (polydioxanone) monofilament suture
Official notice
25 Nov 2022
An official Health Canada notice concerns Pds (polydioxanone) monofilament suture. Returned representative unopened samples were evaluated for knot tensile strength and a portion of samples were observed to be significantly below the minimum knot tensile force requirement. Low suture tensile strength may result in poor performance of the impacted product. The intended benefit of tissue approximation and/or ligation may not be achieved or sustained. Because the knot tensile strength was observed to be significantly below the force requirement, there is a high likelihood that if this issue is present, it will result in breakage prior to use or intra-operatively rather than premature post-operative breakage. Recall start date: November 15, 2022 Published 2022-11-25.
What to do now
Contact the manufacturer if you require additional information.
Recall number
72216
Official notice
2022-11-25
The hazard
The hazard
Returned representative unopened samples were evaluated for knot tensile strength and a portion of samples were observed to be significantly below the minimum knot tensile force requirement. Low suture tensile strength may result in poor performance of the impacted product. The intended benefit of tissue approximation and/or ligation may not be achieved or sustained. Because the knot tensile strength was observed to be significantly below the force requirement, there is a high likelihood that if this issue is present, it will result in breakage prior to use or intra-operatively rather than premature post-operative breakage. Recall start date: November 15, 2022
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Ethicon, Pds Monofilament Suture.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.