No image published for this recall

No image published for this recall

Pds (polydioxanone) monofilament suture

Official notice

25 Nov 2022

Last checked

25 Nov 2022

Medium riskHealth Canada

An official Health Canada notice concerns Pds (polydioxanone) monofilament suture. Returned representative unopened samples were evaluated for knot tensile strength and a portion of samples were observed to be significantly below the minimum knot tensile force requirement. Low suture tensile strength may result in poor performance of the impacted product. The intended benefit of tissue approximation and/or ligation may not be achieved or sustained. Because the knot tensile strength was observed to be significantly below the force requirement, there is a high likelihood that if this issue is present, it will result in breakage prior to use or intra-operatively rather than premature post-operative breakage. Recall start date: November 15, 2022 Published 2022-11-25.

What to do now

Contact the manufacturer if you require additional information.

Recall number

72216

Official notice

2022-11-25

The hazard

Returned representative unopened samples were evaluated for knot tensile strength and a portion of samples were observed to be significantly below the minimum knot tensile force requirement. Low suture tensile strength may result in poor performance of the impacted product. The intended benefit of tissue approximation and/or ligation may not be achieved or sustained. Because the knot tensile strength was observed to be significantly below the force requirement, there is a high likelihood that if this issue is present, it will result in breakage prior to use or intra-operatively rather than premature post-operative breakage. Recall start date: November 15, 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

72216

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - PDS (Polydioxanone) Monofilament Suture, Lot or serial number - SCMSHK, Model or catalog number - Z340H

Risk / hazard

Returned representative unopened samples were evaluated for knot tensile strength and a portion of samples were observed to be significantly below the minimum knot tensile force requirement. Low suture tensile strength may result in poor performance of the impacted product. The intended benefit of tissue approximation and/or ligation may not be achieved or sustained. Because the knot tensile strength was observed to be significantly below the force requirement, there is a high likelihood that if this issue is present, it will result in breakage prior to use or intra-operatively rather than premature post-operative breakage. Recall start date: November 15, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Ethicon, Pds Monofilament Suture.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

72216

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 25 Nov 2022. The next one won't wait for you to search.