Open official notice

Interstim™ systems

Official notice

24 Sept 2026

Last checked

13 Nov 2026

RecallScope risk estimate:HighHealth Canada

An official Health Canada notice concerns Interstim™ systems. Medtronic has received 15 complaints involving the selection of an incompatible component based on the intended system configuration between 01-Jan-2020 through 12-Aug-2026. Use of an incompatible lead or percutaneous extension has resulted in a delay of surgery or the need for surgical replacement of the component. Review of complaint data confirmed that the product is functioning as designed and there are no device or product performance issues. Published 2026-09-24.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82674

Official notice

2026-09-24

The hazard

Medtronic has received 15 complaints involving the selection of an incompatible component based on the intended system configuration between 01-Jan-2020 through 12-Aug-2026. Use of an incompatible lead or percutaneous extension has resulted in a delay of surgery or the need for surgical replacement of the component. Review of complaint data confirmed that the product is functioning as designed and there are no device or product performance issues.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

3560030, 3560022, 978B133, 978B141, 978A128, 978A133, 978B128

Other

RA RA-82674

Model numbers

3560030, 3560022, 978B133, 978B141, 978A128, 978A133, 978B128

Recall Numbers

82674

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 3560030

Product 2

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 3560022

Product 3

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 978B133

Product 4

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 978B141

Product 5

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 978A128

Product 6

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 978A133

Product 7

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 978B128

Hazard

Medtronic has received 15 complaints involving the selection of an incompatible component based on the intended system configuration between 01-Jan-2020 through 12-Aug-2026. Use of an incompatible lead or percutaneous extension has resulted in a delay of surgery or the need for surgical replacement of the component. Review of complaint data confirmed that the product is functioning as designed and there are no device or product performance issues.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Medtronic.

Source & status

Tier

beta

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

82674

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 24 Sept 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button