Open official notice

Architect i system - i2000sr processing module - class 4

Official notice

16 Oct 2026

Last checked

13 Nov 2026

RecallScope risk estimate:HighHealth Canada

An official Health Canada notice concerns Architect i system - i2000sr processing module - class 4. An issue has been identified with the 5 Amp (5A) diode used in the Induction Heater (IH) Control Module on the ARCHITECT i2000SR IH Kit (optional hardware). The diode may be exposed to current exceeding its specified rating, potentially resulting in component failure. Failure could cause localized electrical arcing, circuit board damage, and smoke generation. Only ARCHITECT i2000SR Processing Modules with an Induction Heater (IH) Control Module are impacted.The IH Control Module housing is classified as a fire enclosure in accordance with International Electrotechnical Commission (IEC) 61010-1 and is designed to contain thermal events. Protective features include flame-retardant circuit board material and a metal fire-rated enclosure, which help prevent propagation beyond the module. In all failures, the observed damage remained contained within the IH Control Module, and no injuries or property damage have been reported.There is potential for operator injury and property damage due to smoke, burn, hot surfaces, etc. Caused by the excess current on the diode. There have been no occurrences of fire and smoke leading to property damage and no reported occurrences of operator injury for this issue. Published 2026-10-16.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82800

Official notice

2026-10-16

The hazard

An issue has been identified with the 5 Amp (5A) diode used in the Induction Heater (IH) Control Module on the ARCHITECT i2000SR IH Kit (optional hardware). The diode may be exposed to current exceeding its specified rating, potentially resulting in component failure. Failure could cause localized electrical arcing, circuit board damage, and smoke generation. Only ARCHITECT i2000SR Processing Modules with an Induction Heater (IH) Control Module are impacted.The IH Control Module housing is classified as a fire enclosure in accordance with International Electrotechnical Commission (IEC) 61010-1 and is designed to contain thermal events. Protective features include flame-retardant circuit board material and a metal fire-rated enclosure, which help prevent propagation beyond the module. In all failures, the observed damage remained contained within the IH Control Module, and no injuries or property damage have been reported.There is potential for operator injury and property damage due to smoke, burn, hot surfaces, etc. Caused by the excess current on the diode. There have been no occurrences of fire and smoke leading to property damage and no reported occurrences of operator injury for this issue.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

ISR64041

Skus

03M74-02

Other

RA RA-82800

Model numbers

03M74-02

Recall Numbers

82800

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - ISR64041, Model or catalogue number - 03M74-02

Hazard

An issue has been identified with the 5 Amp (5A) diode used in the Induction Heater (IH) Control Module on the ARCHITECT i2000SR IH Kit (optional hardware). The diode may be exposed to current exceeding its specified rating, potentially resulting in component failure. Failure could cause localized electrical arcing, circuit board damage, and smoke generation. Only ARCHITECT i2000SR Processing Modules with an Induction Heater (IH) Control Module are impacted.The IH Control Module housing is classified as a fire enclosure in accordance with International Electrotechnical Commission (IEC) 61010-1 and is designed to contain thermal events. Protective features include flame-retardant circuit board material and a metal fire-rated enclosure, which help prevent propagation beyond the module. In all failures, the observed damage remained contained within the IH Control Module, and no injuries or property damage have been reported.There is potential for operator injury and property damage due to smoke, burn, hot surfaces, etc. Caused by the excess current on the diode. There have been no occurrences of fire and smoke leading to property damage and no reported occurrences of operator injury for this issue.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Abbott Laboratories.

Source & status

Tier

beta

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

82800

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 16 Oct 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button