Alinity hs hq system
Official notice
23 Sept 2022
An official Health Canada notice concerns Alinity hs hq system. The Alinity h-series system control center (SCC) application software version 4.3 (and below) and the local user interface (LUI) embedded analyzer engine (EAE) contain a cybersecurity certificate that expired on 08sep2022. Whenever the software checks for the cybersecurity certificate expiration date and it detects that it has expired, the SCC and/or LUI becomes nonfunctional. This renders the Alinity h-series unable to process samples. There is no impact to cybersecurity or results integrity on the Alinity h-series with this issue. Recall start date: September 12, 2022 Published 2022-09-23.
What to do now
Contact the manufacturer if you require additional information.
Recall number
64565
Official notice
2022-09-23
The hazard
The hazard
The Alinity h-series system control center (SCC) application software version 4.3 (and below) and the local user interface (LUI) embedded analyzer engine (EAE) contain a cybersecurity certificate that expired on 08sep2022. Whenever the software checks for the cybersecurity certificate expiration date and it detects that it has expired, the SCC and/or LUI becomes nonfunctional. This renders the Alinity h-series unable to process samples. There is no impact to cybersecurity or results integrity on the Alinity h-series with this issue. Recall start date: September 12, 2022
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Abbott Laboratories, Alinity Hs Hq System.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.