Alinity hq analyzer
Official notice
2 Oct 2025
An official Health Canada notice concerns Alinity hq analyzer. An issue has been identified in Alinity HQ Analyzer software versions 5.0 to 5.10.1. The immature granulocyte saturation (IG sat) flag is intended to indicate a high number of saturated events that impact the IG/%IG and neutrophils (NEU/%NEU) parameters. However, the system does not visibly apply appropriate flagging to IG/%IG and NEU/%NEU results when saturation is present, which may prevent users from recognizing potentially impacted values. When IG sat is present, NEU is the most likely to be affected. Additionally, because white blood cell (WBC) differential results must total 100%, any distortion in IG values may also affect other parameters-such as basophils (BASO), monocytes (MONO), lymphocytes (LYM), and eosinophils (EOS) -potentially leading to inaccurate distribution across the full wbc differential. A new software version for the Alinity HQ analyzer to resolve this issue is expected to be available in 2026. Your Abbott representative will contact you to schedule a mandatory upgrade. Published 2025-10-02.
What to do now
Contact the manufacturer if you require additional information.
Recall number
78175
Official notice
2025-10-02
The hazard
The hazard
An issue has been identified in Alinity HQ Analyzer software versions 5.0 to 5.10.1. The immature granulocyte saturation (IG sat) flag is intended to indicate a high number of saturated events that impact the IG/%IG and neutrophils (NEU/%NEU) parameters. However, the system does not visibly apply appropriate flagging to IG/%IG and NEU/%NEU results when saturation is present, which may prevent users from recognizing potentially impacted values. When IG sat is present, NEU is the most likely to be affected. Additionally, because white blood cell (WBC) differential results must total 100%, any distortion in IG values may also affect other parameters-such as basophils (BASO), monocytes (MONO), lymphocytes (LYM), and eosinophils (EOS) -potentially leading to inaccurate distribution across the full wbc differential. A new software version for the Alinity HQ analyzer to resolve this issue is expected to be available in 2026. Your Abbott representative will contact you to schedule a mandatory upgrade.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Abbott Laboratories, Alinity Hq Analyzer.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.