Dxc 500i clinical analyzer and dxc 500 au clinical chemistry analyzer

Official notice

10 Sept 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Dxc 500i clinical analyzer and dxc 500 au clinical chemistry analyzer. Beckman Coulter identified a software issue affecting DxC 500i and DxC 500 AU Analyzers when testing plasma samples. Following user configuration of a chemistry test to nondefault units, subsequent activation of a draft configuration in the system configuration page may revert units to the system default without notifying the user. Published 2026-09-10.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82613

Official notice

2026-09-10

The hazard

Beckman Coulter identified a software issue affecting DxC 500i and DxC 500 AU Analyzers when testing plasma samples. Following user configuration of a chemistry test to nondefault units, subsequent activation of a draft configuration in the system configuration page may revert units to the system default without notifying the user.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

C63521, C63522, C63520, C63519

Other

RA RA-82613

Model numbers

C63521, C63522, C63520, C63519

Recall Numbers

82613

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - Software V. 1.3.5 or earlier, Model or catalogue number - C63521

Product 2

Lot or serial number - Software V. 1.3.5 or earlier, Model or catalogue number - C63522

Product 3

Lot or serial number - Software V. 1.4.5 or earlier, Model or catalogue number - C63520

Product 4

Lot or serial number - Software V. 1.4.5 or earlier, Model or catalogue number - C63519

Hazard

Beckman Coulter identified a software issue affecting DxC 500i and DxC 500 AU Analyzers when testing plasma samples. Following user configuration of a chemistry test to nondefault units, subsequent activation of a draft configuration in the system configuration page may revert units to the system default without notifying the user.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Beckman Coulter.

Source & status

Tier

beta

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

82613

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 10 Sept 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button