Unicel dxh series instruments

Official notice

20 Aug 2025

Last checked

20 Aug 2025

Medium riskHealth Canada
Watch Beckman Coulter

An official Health Canada notice concerns Unicel dxh series instruments. Beckman Coulter has identified a potential for erroneously high hemoglobin (Hgb) results on samples with elevated white blood cell (WBC) counts. This occurs on Unicel DxH series instruments configured with the led Hgb photometer. Falsely elevated Hgb results may observed in specimens with WBC counts > 30 x 10³ cells/μl. Recall start date: August 6, 2025 Published 2025-08-20.

What to do now

Contact the manufacturer if you require additional information.

Recall number

77942

Official notice

2025-08-20

The hazard

Beckman Coulter has identified a potential for erroneously high hemoglobin (Hgb) results on samples with elevated white blood cell (WBC) counts. This occurs on Unicel DxH series instruments configured with the led Hgb photometer. Falsely elevated Hgb results may observed in specimens with WBC counts > 30 x 10³ cells/μl. Recall start date: August 6, 2025

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Recall Numbers

77942

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Affected products - Unicel DxH 900 Coulter Cellular Analysis System, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - C11478

Product 2

Affected products - Unicel DxH 600 Coulter Cellular Analysis System, Lot or serial number - BB49881, Model or catalogue number - B23858

Product 3

Affected products - Unicel DxH 800 Coulter Cellular Analysis System, Lot or serial number - AU10078, Model or catalogue number - 629029

Product 4

Affected products - Unicel DxH 800 Coulter Cellular Analysis System, Lot or serial number - AW43276, Model or catalogue number - 629029

Product 5

Affected products - Unicel DxH 800 Coulter Cellular Analysis System, Lot or serial number - AZ46681, Model or catalogue number - 629029

Product 6

Affected products - Unicel DxH 800 Coulter Cellular Analysis System, Lot or serial number - AU10078, Model or catalogue number - B24802

Product 7

Affected products - Unicel DxH 800 Coulter Cellular Analysis System, Lot or serial number - AW43276, Model or catalogue number - B24802

Product 8

Affected products - Unicel DxH 800 Coulter Cellular Analysis System, Lot or serial number - AZ46681, Model or catalogue number - B24802

Product 9

Affected products - Unicel DxH 690T Coulter Cellular Analysis System, Lot or serial number - BJ12047, Model or catalogue number - C34520

Product 10

Affected products - Unicel DxH 690T Coulter Cellular Analysis System, Lot or serial number - BJ07013, Model or catalogue number - C34520

Product 11

Affected products - Unicel DxH 690T Coulter Cellular Analysis System, Lot or serial number - BJ12046, Model or catalogue number - C34520

Product 12

Affected products - Unicel DxH 690T Coulter Cellular Analysis System, Lot or serial number - BH24076, Model or catalogue number - C34520

Risk / hazard

Beckman Coulter has identified a potential for erroneously high hemoglobin (Hgb) results on samples with elevated white blood cell (WBC) counts. This occurs on Unicel DxH series instruments configured with the led Hgb photometer. Falsely elevated Hgb results may observed in specimens with WBC counts > 30 x 10³ cells/μl. Recall start date: August 6, 2025

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Beckman Coulter, Unicel Dxh Series Instruments

Manufacturer

Not provided

Phone

Not provided

Brand: Beckman Coulter, Unicel Dxh Series Instruments.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

77942

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Unicel dxi 600 access immunoassay system-total t4 assay

Medium riskHealth Canada19 Mar 2026
CA
Recall image placeholder

Dxc 700 au and au5800 clinical chemistry analyzers

Medium riskHealth Canada5 Feb 2026
CA
Recall image placeholder

Cellmek sps

Medium riskHealth Canada30 Dec 2025
CA
Recall image placeholder

Unicel dxi 600 and 800 access immunoassay systems

Medium riskHealth Canada16 Dec 2025
CA
Recall image placeholder

Beckman coulter au immunochemistry system - bicarbonate (co2) assay

Medium riskHealth Canada17 Oct 2025
CA
Recall image placeholder

Dxc 500i clinical analyzer and dxi 9000 access immunoassay analyzer

Critical riskHealth Canada26 Sept 2025
CA
Recall image placeholder

Synchron® systems direct ldl cholesterol (ldld) reagent

Medium riskHealth Canada25 Sept 2025
CA
Recall image placeholder

Synchron systems phosphorus (phosm) reagent

Medium riskHealth Canada17 Sept 2025
CA
Recall image placeholder

Cellmek sps

Medium riskHealth Canada4 Sept 2025
CA
Recall image placeholder

Dxc 500i clinical analyzers

Medium riskHealth Canada19 Aug 2025
CA
Recall image placeholder

Unicel dxi 600 access immunoassay system- thyroglobulin assay

Medium riskHealth Canada23 Jul 2025
CA
Recall image placeholder

Unicel dxi 600 & 800 systems

Medium riskHealth Canada16 Jul 2025

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 20 Aug 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.