Unicel dxh series instruments

Official notice

20 Aug 2025

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Unicel dxh series instruments. Beckman Coulter has identified a potential for erroneously high hemoglobin (Hgb) results on samples with elevated white blood cell (WBC) counts. This occurs on Unicel DxH series instruments configured with the led Hgb photometer. Falsely elevated Hgb results may observed in specimens with WBC counts > 30 x 10³ cells/μl. Recall start date: August 6, 2025 Published 2025-08-20.

What to do now

Contact the manufacturer if you require additional information.

Recall number

77942

Official notice

2025-08-20

The hazard

Beckman Coulter has identified a potential for erroneously high hemoglobin (Hgb) results on samples with elevated white blood cell (WBC) counts. This occurs on Unicel DxH series instruments configured with the led Hgb photometer. Falsely elevated Hgb results may observed in specimens with WBC counts > 30 x 10³ cells/μl. Recall start date: August 6, 2025

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

77942

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - Unicel DxH 900 Coulter Cellular Analysis System, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - C11478

Product 2

Affected products - Unicel DxH 600 Coulter Cellular Analysis System, Lot or serial number - BB49881, Model or catalogue number - B23858

Product 3

Affected products - Unicel DxH 800 Coulter Cellular Analysis System, Lot or serial number - AU10078, Model or catalogue number - 629029

Product 4

Affected products - Unicel DxH 800 Coulter Cellular Analysis System, Lot or serial number - AW43276, Model or catalogue number - 629029

Product 5

Affected products - Unicel DxH 800 Coulter Cellular Analysis System, Lot or serial number - AZ46681, Model or catalogue number - 629029

Product 6

Affected products - Unicel DxH 800 Coulter Cellular Analysis System, Lot or serial number - AU10078, Model or catalogue number - B24802

Product 7

Affected products - Unicel DxH 800 Coulter Cellular Analysis System, Lot or serial number - AW43276, Model or catalogue number - B24802

Product 8

Affected products - Unicel DxH 800 Coulter Cellular Analysis System, Lot or serial number - AZ46681, Model or catalogue number - B24802

Product 9

Affected products - Unicel DxH 690T Coulter Cellular Analysis System, Lot or serial number - BJ12047, Model or catalogue number - C34520

Product 10

Affected products - Unicel DxH 690T Coulter Cellular Analysis System, Lot or serial number - BJ07013, Model or catalogue number - C34520

Product 11

Affected products - Unicel DxH 690T Coulter Cellular Analysis System, Lot or serial number - BJ12046, Model or catalogue number - C34520

Product 12

Affected products - Unicel DxH 690T Coulter Cellular Analysis System, Lot or serial number - BH24076, Model or catalogue number - C34520

Hazard

Beckman Coulter has identified a potential for erroneously high hemoglobin (Hgb) results on samples with elevated white blood cell (WBC) counts. This occurs on Unicel DxH series instruments configured with the led Hgb photometer. Falsely elevated Hgb results may observed in specimens with WBC counts > 30 x 10³ cells/μl. Recall start date: August 6, 2025

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Beckman Coulter, Unicel Dxh Series Instruments.

Source & status

Tier

official

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

77942

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 20 Aug 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button