Unicel dxh series instruments
Official notice
20 Aug 2025
An official Health Canada notice concerns Unicel dxh series instruments. Beckman Coulter has identified a potential for erroneously high hemoglobin (Hgb) results on samples with elevated white blood cell (WBC) counts. This occurs on Unicel DxH series instruments configured with the led Hgb photometer. Falsely elevated Hgb results may observed in specimens with WBC counts > 30 x 10³ cells/μl. Recall start date: August 6, 2025 Published 2025-08-20.
What to do now
Contact the manufacturer if you require additional information.
Recall number
77942
Official notice
2025-08-20
The hazard
The hazard
Beckman Coulter has identified a potential for erroneously high hemoglobin (Hgb) results on samples with elevated white blood cell (WBC) counts. This occurs on Unicel DxH series instruments configured with the led Hgb photometer. Falsely elevated Hgb results may observed in specimens with WBC counts > 30 x 10³ cells/μl. Recall start date: August 6, 2025
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Product 2
Product 3
Product 4
Product 5
Product 6
Product 7
Product 8
Product 9
Product 10
Product 11
Product 12
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Beckman Coulter, Unicel Dxh Series Instruments.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.