Dxc 500i clinical analyzer and dxi 9000 access immunoassay analyzer

Official notice

26 Sept 2025

RecallScope risk estimate:CriticalHealth Canada

An official Health Canada notice concerns Dxc 500i clinical analyzer and dxi 9000 access immunoassay analyzer. Beckman Coulter has confirmed a software anomaly on the DxC 500i Clinical Analyzer and DxI 9000 Access Immunoassay Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results (R1, R2) and potential delays in reporting (R3), risking erroneous diagnoses if undetected Published 2025-09-26.

What to do now

Contact the manufacturer if you require additional information.

Recall number

78147

Official notice

2025-09-26

The hazard

Beckman Coulter has confirmed a software anomaly on the DxC 500i Clinical Analyzer and DxI 9000 Access Immunoassay Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results (R1, R2) and potential delays in reporting (R3), risking erroneous diagnoses if undetected

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

78147

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - DxI 9000 Access Immunoassay Analyzer, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - C13252

Hazard

Beckman Coulter has confirmed a software anomaly on the DxC 500i Clinical Analyzer and DxI 9000 Access Immunoassay Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results (R1, R2) and potential delays in reporting (R3), risking erroneous diagnoses if undetected

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Beckman Coulter, Dxc 500i Clinical Analyzer And Dxi 9000 Access Immunoassay Analyzer.

Source & status

Tier

official

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

78147

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 26 Sept 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button