19 Mar 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Unicel dxi 600 access immunoassay system-total t4 assay. Access Total T4 calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. In the worst-case scenario, there is a medium safety risk associated with this issue due to the potential to cause false low patient results. Published 2026-03-19.

What to do now

Contact the manufacturer if you require additional information.

Recall number

81777

Official notice

2026-03-19

The hazard

Access Total T4 calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. In the worst-case scenario, there is a medium safety risk associated with this issue due to the potential to cause false low patient results.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

538815, 539243, 538729, 538898, 539107

Skus

33805

Other

RA RA-81777

Model numbers

33805

Recall Numbers

81777

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - 538815, Model or catalogue number - 33805

Product 2

Lot or serial number - 539243, Model or catalogue number - 33805

Product 3

Lot or serial number - 538729, Model or catalogue number - 33805

Product 4

Lot or serial number - 538898, Model or catalogue number - 33805

Product 5

Lot or serial number - 539107, Model or catalogue number - 33805

Hazard

Access Total T4 calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. In the worst-case scenario, there is a medium safety risk associated with this issue due to the potential to cause false low patient results.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Beckman Coulter.

Source & status

Tier

beta

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

81777

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 19 Mar 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button