Open official notice

Dxc 500i clinical analyzers

Official notice

19 Aug 2025

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Dxc 500i clinical analyzers. Beckman Coulter has determined that on the DxC 500i clinical analyzer, when ordering a clinical chemistry (CC) combination test simultaneously with an immunoassay (IA) test, if the IA test is processed between the CC constituent tests, the sample status of the CC test will remain "in progress" with no error events. Recall start date: August 4, 2025 Published 2025-08-19.

What to do now

Contact the manufacturer if you require additional information.

Recall number

77931

Official notice

2025-08-19

The hazard

Beckman Coulter has determined that on the DxC 500i clinical analyzer, when ordering a clinical chemistry (CC) combination test simultaneously with an immunoassay (IA) test, if the IA test is processed between the CC constituent tests, the sample status of the CC test will remain "in progress" with no error events. Recall start date: August 4, 2025

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

77931

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - DxC 500i Clinical Analyzer, DxC 500i Au Chemistry Module, Lot or serial number - software 1.3.3 or 1.3.4, Model or catalogue number - C63522

Product 2

Affected products - DxC 500i Clinical Analyzer, DxC 500i Access 2 Immunoassay Module, Lot or serial number - software 1.3.3 or 1.3.4, Model or catalogue number - C13252

Hazard

Beckman Coulter has determined that on the DxC 500i clinical analyzer, when ordering a clinical chemistry (CC) combination test simultaneously with an immunoassay (IA) test, if the IA test is processed between the CC constituent tests, the sample status of the CC test will remain "in progress" with no error events. Recall start date: August 4, 2025

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Beckman Coulter, Dxc 500i Clinical Analyzers.

Source & status

Tier

official

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

77931

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 19 Aug 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.