Open official notice

Automated blood coagulation analyzers

Official notice

6 Nov 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Automated blood coagulation analyzers. In October 2026, the legal manufacturer Sysmex notified Siemens Healthineers of an urgent field safety notice for the analyzer CN series systems.Sysmex identified that deposits may accumulate on the sample arm B probe of the CN-series analyzers.The deposits may affect sample dispensing, potentially resulting in the delivery of a lower sample volume to the reaction cuvette than defined, which and may cause falsely decreased results. Based on the manufacturer’s investigation, the accumulation of deposits occurred due to insufficient washing of the sample arm B probe independent from the assays used. Published 2026-11-06.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82927

Official notice

2026-11-06

The hazard

In October 2026, the legal manufacturer Sysmex notified Siemens Healthineers of an urgent field safety notice for the analyzer CN series systems.Sysmex identified that deposits may accumulate on the sample arm B probe of the CN-series analyzers.The deposits may affect sample dispensing, potentially resulting in the delivery of a lower sample volume to the reaction cuvette than defined, which and may cause falsely decreased results. Based on the manufacturer’s investigation, the accumulation of deposits occurred due to insufficient washing of the sample arm B probe independent from the assays used.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

Q1079, Q1078

Skus

BN343315, CL971029, AX801645, CV339763

Other

RA RA-82927

Model numbers

BN343315, CL971029, AX801645, CV339763

Recall Numbers

82927

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - BN343315

Product 2

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - CL971029

Product 3

Lot or serial number - Q1079, Model or catalogue number - AX801645

Product 4

Lot or serial number - Q1078, Model or catalogue number - AX801645

Product 5

Lot or serial number - Q1079, Model or catalogue number - CV339763

Product 6

Lot or serial number - Q1078, Model or catalogue number - CV339763

Hazard

In October 2026, the legal manufacturer Sysmex notified Siemens Healthineers of an urgent field safety notice for the analyzer CN series systems.Sysmex identified that deposits may accumulate on the sample arm B probe of the CN-series analyzers.The deposits may affect sample dispensing, potentially resulting in the delivery of a lower sample volume to the reaction cuvette than defined, which and may cause falsely decreased results. Based on the manufacturer’s investigation, the accumulation of deposits occurred due to insufficient washing of the sample arm B probe independent from the assays used.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Manufacturer

Not provided

Phone

Not provided

Brand: Sysmex.

Source & status

Tier

beta

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

82927

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 6 Nov 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button