Iq200 series analyzers and dxu 840m 850m iris analyzers

Official notice

21 Aug 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Iq200 series analyzers and dxu 840m 850m iris analyzers. Beckman Coulter received customer complaints of lamina fluid leaks in iQ200 Series and DxU 840m/850m Analyzers. The investigation found that a structurally compromised rinse-line check valve may allow fluid to migrate internally and leak onto the table or floor, potentially resulting into exposure to physical or mechanical or chemical hazards. Published 2026-08-21.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82520

Official notice

2026-08-21

The hazard

Beckman Coulter received customer complaints of lamina fluid leaks in iQ200 Series and DxU 840m/850m Analyzers. The investigation found that a structurally compromised rinse-line check valve may allow fluid to migrate internally and leak onto the table or floor, potentially resulting into exposure to physical or mechanical or chemical hazards.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

C76947, 700-3325, 700-3345, C49513, 700-3375

Other

RA RA-82520

Model numbers

C76947, 700-3325, 700-3345, C49513, 700-3375

Recall Numbers

82520

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - C76947

Product 2

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 700-3325

Product 3

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 700-3345

Product 4

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - C49513

Product 5

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 700-3375

Hazard

Beckman Coulter received customer complaints of lamina fluid leaks in iQ200 Series and DxU 840m/850m Analyzers. The investigation found that a structurally compromised rinse-line check valve may allow fluid to migrate internally and leak onto the table or floor, potentially resulting into exposure to physical or mechanical or chemical hazards.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Beckman Coulter Ireland.

Source & status

Tier

beta

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

82520

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 21 Aug 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button