Open official notice

Irispec ca cb cc

Official notice

14 Mar 2022

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Irispec ca cb cc. Beckman Coulter has become aware of an increase in complaints generated for Irispec Ca failures for glucose on the ichemvelocity instrument. If a Qc run fails, the system will not allow the operator to run specimens. The Qc run should be repeated and be successful before running patient specimens. This issue affects Irispec Ca control lot number 583-21 and higher when analyzed with ichemvelocity urine chemistry strips lot number 7212510m and higher. There is no impact to patient results, Irispec Cb or Cc controls. Recall start date: Feb 14, 2022 Published 2022-03-14.

What to do now

Contact the manufacturer if you require additional information.

Recall number

63984

Official notice

2022-03-14

The hazard

Beckman Coulter has become aware of an increase in complaints generated for Irispec Ca failures for glucose on the ichemvelocity instrument. If a Qc run fails, the system will not allow the operator to run specimens. The Qc run should be repeated and be successful before running patient specimens. This issue affects Irispec Ca control lot number 583-21 and higher when analyzed with ichemvelocity urine chemistry strips lot number 7212510m and higher. There is no impact to patient results, Irispec Cb or Cc controls. Recall start date: Feb 14, 2022

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

63984

Lots

583-21, 7212510m

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected Products - Irispec Ca/Cb/Cc, Lot or Serial Number - CA control 583-21 and higher, Model or catalog number - 800-7211

Hazard

Beckman Coulter has become aware of an increase in complaints generated for Irispec Ca failures for glucose on the ichemvelocity instrument. If a Qc run fails, the system will not allow the operator to run specimens. The Qc run should be repeated and be successful before running patient specimens. This issue affects Irispec Ca control lot number 583-21 and higher when analyzed with ichemvelocity urine chemistry strips lot number 7212510m and higher. There is no impact to patient results, Irispec Cb or Cc controls. Recall start date: Feb 14, 2022

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Beckman Coulter Ireland, Irispec Ca Cb Cc.

Source & status

Tier

official

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

63984

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 14 Mar 2022. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button