No image published for this recall
Irispec ca cb cc
Official notice
14 Mar 2022
Last checked
14 Mar 2022
An official Health Canada notice concerns Irispec ca cb cc. Beckman Coulter has become aware of an increase in complaints generated for Irispec Ca failures for glucose on the ichemvelocity instrument. If a Qc run fails, the system will not allow the operator to run specimens. The Qc run should be repeated and be successful before running patient specimens. This issue affects Irispec Ca control lot number 583-21 and higher when analyzed with ichemvelocity urine chemistry strips lot number 7212510m and higher. There is no impact to patient results, Irispec Cb or Cc controls. Recall start date: Feb 14, 2022 Published 2022-03-14.
What to do now
Contact the manufacturer if you require additional information.
Recall number
63984
Official notice
2022-03-14
The hazard
The hazard
Beckman Coulter has become aware of an increase in complaints generated for Irispec Ca failures for glucose on the ichemvelocity instrument. If a Qc run fails, the system will not allow the operator to run specimens. The Qc run should be repeated and be successful before running patient specimens. This issue affects Irispec Ca control lot number 583-21 and higher when analyzed with ichemvelocity urine chemistry strips lot number 7212510m and higher. There is no impact to patient results, Irispec Cb or Cc controls. Recall start date: Feb 14, 2022
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Lots
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Beckman Coulter Ireland, Irispec Ca Cb Cc.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Irispec ca cb cc

Irispec ca cb cc

Iricell, iq 200 and dxu urine analyzers

Iq 200 and dxu systems

Ortho optix™ reader

Irispec ca cb cc

Solar™ gi

Rapidlab® 348ex system

Nephromax high pressure nephrostomy balloon catheter

Uhi-4 high flow insufflation system - insufflator

Atellica ch enzymatic hemoglobin a1c (a1c e) assay

Gsp® neonatal galt kit
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 14 Mar 2022. The next one won't wait for you to search.