Irispec ca cb cc
Official notice
14 Mar 2022
An official Health Canada notice concerns Irispec ca cb cc. Beckman Coulter has become aware of an increase in complaints generated for Irispec Ca failures for glucose on the ichemvelocity instrument. If a Qc run fails, the system will not allow the operator to run specimens. The Qc run should be repeated and be successful before running patient specimens. This issue affects Irispec Ca control lot number 583-21 and higher when analyzed with ichemvelocity urine chemistry strips lot number 7212510m and higher. There is no impact to patient results, Irispec Cb or Cc controls. Recall start date: Feb 14, 2022 Published 2022-03-14.
What to do now
Contact the manufacturer if you require additional information.
Recall number
63984
Official notice
2022-03-14
The hazard
The hazard
Beckman Coulter has become aware of an increase in complaints generated for Irispec Ca failures for glucose on the ichemvelocity instrument. If a Qc run fails, the system will not allow the operator to run specimens. The Qc run should be repeated and be successful before running patient specimens. This issue affects Irispec Ca control lot number 583-21 and higher when analyzed with ichemvelocity urine chemistry strips lot number 7212510m and higher. There is no impact to patient results, Irispec Cb or Cc controls. Recall start date: Feb 14, 2022
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Recall Numbers
Lots
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Beckman Coulter Ireland, Irispec Ca Cb Cc.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.