Open official notice

Iricell, iq 200 and dxu urine analyzers

Official notice

7 Feb 2024

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Iricell, iq 200 and dxu urine analyzers. Beckman Coulter has become aware that the cannula, which is found in the flowcell and the IQclear rinse waste well assembly, may rust. Rust from the cannula has the potential to generate erroneous results. Recall start date: December 18, 2023 Published 2024-02-07.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75031

Official notice

2024-02-07

The hazard

Beckman Coulter has become aware that the cannula, which is found in the flowcell and the IQclear rinse waste well assembly, may rust. Rust from the cannula has the potential to generate erroneous results. Recall start date: December 18, 2023

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

75031

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - iRICELL 2000 Plus, Lot or serial number - 7871, Model or catalog number - 800-7714

Product 2

Affected products - iRICELL 3000 Plus, Lot or serial number - 7871, Model or catalog number - 800-3079 800-7713

Product 3

Affected products - IQ 200 Sprint Plus W Load/Unload, Lot or serial number - 7871, Model or catalog number - 800-3081

Product 4

Affected products - iRICELL 2000 Pro, Lot or serial number - 7871, Model or catalog number - 800-3049

Product 5

Affected products - DxU 850m/840m Microscopy Analyzers, Lot or serial number - 100188, Model or catalog number - C76947

Product 6

Affected products - IQ 200 Sprint Pro, Lot or serial number - 7871, Model or catalog number - 800-3053

Product 7

Affected products - IQ 200 Sprint Pro W Load/Unload, Lot or serial number - 7871, Model or catalog number - 800-3082

Product 8

Affected products - IQ 200 Sprint Plus, Lot or serial number - 7871, Model or catalog number - 800-3933

Product 9

Affected products - iRICELL3000 Pro, Lot or serial number - 7871, Model or catalog number - 800-3080 800-3050

Hazard

Beckman Coulter has become aware that the cannula, which is found in the flowcell and the IQclear rinse waste well assembly, may rust. Rust from the cannula has the potential to generate erroneous results. Recall start date: December 18, 2023

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Beckman Coulter Ireland, Iricell Iq 200 And Dxu Urine Analyzers.

Source & status

Tier

official

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

75031

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 7 Feb 2024. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button