Irispec ca cb cc

Official notice

26 Jul 2024

Last checked

26 Jul 2024

Medium riskHealth Canada

An official Health Canada notice concerns Irispec ca cb cc. Beckman Coulter has become aware of an increase in complaints for intermittent IRISpec CB failures for ketone on the iChemVELOCITY Urine Chemistry System (instructions for use PN 300-4449FH) and DxU 810c IRIS Urine Chemistry System (instructions for use PN C49355AA). If a QC run fails, the system will not allow the operator to run patient samples. The QC run should be repeated and pass successfully before running patient samples. Recall start date: July 16, 2024 Published 2024-07-26.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75875

Official notice

2024-07-26

The hazard

Beckman Coulter has become aware of an increase in complaints for intermittent IRISpec CB failures for ketone on the iChemVELOCITY Urine Chemistry System (instructions for use PN 300-4449FH) and DxU 810c IRIS Urine Chemistry System (instructions for use PN C49355AA). If a QC run fails, the system will not allow the operator to run patient samples. The QC run should be repeated and pass successfully before running patient samples. Recall start date: July 16, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75875

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - IRISpec CA/CB/CC, Lot or serial number - 432-24 and higher, Model or catalog number - 800-7702

Risk / hazard

Beckman Coulter has become aware of an increase in complaints for intermittent IRISpec CB failures for ketone on the iChemVELOCITY Urine Chemistry System (instructions for use PN 300-4449FH) and DxU 810c IRIS Urine Chemistry System (instructions for use PN C49355AA). If a QC run fails, the system will not allow the operator to run patient samples. The QC run should be repeated and pass successfully before running patient samples. Recall start date: July 16, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Beckman Coulter Ireland, Irispec Ca Cb Cc.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

75875

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
No image published for this recall

Irispec ca cb cc

Medium riskHealth Canada26 Jul 2024
CA
No image published for this recall

Iricell, iq 200 and dxu urine analyzers

Medium riskHealth Canada7 Feb 2024
CA
No image published for this recall

Iq 200 and dxu systems

Medium riskHealth Canada3 May 2023
CA
No image published for this recall

Ortho optix™ reader

Medium riskHealth Canada25 Nov 2022
CA
No image published for this recall

Irispec ca cb cc

Medium riskHealth Canada25 Nov 2022
CA
No image published for this recall

Irispec ca cb cc

Medium riskHealth Canada14 Mar 2022
CA
No image published for this recall

Nephromax high pressure nephrostomy balloon catheter

Medium riskHealth Canada2 Jul 2026
CA
No image published for this recall

Uhi-4 high flow insufflation system - insufflator

Medium riskHealth Canada24 Jun 2026
CA
No image published for this recall

Atellica ch enzymatic hemoglobin a1c (a1c e) assay

Medium riskHealth Canada18 Jun 2026
CA
No image published for this recall

Gsp® neonatal galt kit

Medium riskHealth Canada22 May 2026
CA
No image published for this recall

Harmonyair a-series surgical lighting system and alyon surgical lighting system

Medium riskHealth Canada21 May 2026
CA
No image published for this recall

Endoflip system

High riskHealth Canada14 May 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 26 Jul 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.