No image published for this recall

No image published for this recall

Irispec ca cb cc

Official notice

26 Jul 2024

Last checked

26 Jul 2024

Medium riskHealth Canada

An official Health Canada notice concerns Irispec ca cb cc. Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB glucose control failure, resulting in false positive results when used in conjunction with iChemVELOCITY Urine Chemistry Strips PNS 800-7204 and 800- 7212. If a QC run fails, the system will not allow the operator to run patient samples. The QC run should be repeated and pass successfully before running patient samples. Recall start date: July 16, 2024 Published 2024-07-26.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75879

Official notice

2024-07-26

The hazard

Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB glucose control failure, resulting in false positive results when used in conjunction with iChemVELOCITY Urine Chemistry Strips PNS 800-7204 and 800- 7212. If a QC run fails, the system will not allow the operator to run patient samples. The QC run should be repeated and pass successfully before running patient samples. Recall start date: July 16, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75879

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - IRISpec CA/CB/CC, Lot or serial number - 432-24 and higher, Model or catalog number - 800-7211 800-7702

Risk / hazard

Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB glucose control failure, resulting in false positive results when used in conjunction with iChemVELOCITY Urine Chemistry Strips PNS 800-7204 and 800- 7212. If a QC run fails, the system will not allow the operator to run patient samples. The QC run should be repeated and pass successfully before running patient samples. Recall start date: July 16, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Beckman Coulter Ireland, Irispec Ca Cb Cc.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

75879

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 26 Jul 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.