No image published for this recall

Lifecodes lsa class ii

Official notice

2 Jul 2026

MediumHealth Canada

An official Health Canada notice concerns Lifecodes lsa class ii. Werfen has identified that two beads within the LIFECODES LSA class II assay kit(s) may present different DQB1 specificities than previously understood:• Bead 42 / Antigen ID 242Previously associated with DQB1*02:01, this bead instead presents DQB1*02:02• Bead 44 / Antigen ID 244Previously associated with DQB1*02:02, this bead instead presents DQB1*02:01• Note: DQA1*02:01 is present on both beads and therefore not impacted.Due to this issue, certain reactivity patterns may appear differently than expected. Published 2026-07-02.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82270

Official notice

2026-07-02

The hazard

Werfen has identified that two beads within the LIFECODES LSA class II assay kit(s) may present different DQB1 specificities than previously understood:• Bead 42 / Antigen ID 242Previously associated with DQB1*02:01, this bead instead presents DQB1*02:02• Bead 44 / Antigen ID 244Previously associated with DQB1*02:02, this bead instead presents DQB1*02:01• Note: DQA1*02:01 is present on both beads and therefore not impacted.Due to this issue, certain reactivity patterns may appear differently than expected.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Lots

3016850, 3016697, 3016317, 3016532, 3016100

Skus

265200IVD

Other

RA RA-82270

Model Numbers

265200IVD

Recall Numbers

82270

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - 3016850, Model or catalogue number - 265200IVD

Product 2

Lot or serial number - 3016697, Model or catalogue number - 265200IVD

Product 3

Lot or serial number - 3016317, Model or catalogue number - 265200IVD

Product 4

Lot or serial number - 3016532, Model or catalogue number - 265200IVD

Product 5

Lot or serial number - 3016100, Model or catalogue number - 265200IVD

Risk / hazard

Werfen has identified that two beads within the LIFECODES LSA class II assay kit(s) may present different DQB1 specificities than previously understood:• Bead 42 / Antigen ID 242Previously associated with DQB1*02:01, this bead instead presents DQB1*02:02• Bead 44 / Antigen ID 244Previously associated with DQB1*02:02, this bead instead presents DQB1*02:01• Note: DQA1*02:01 is present on both beads and therefore not impacted.Due to this issue, certain reactivity patterns may appear differently than expected.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Immucor Gti Diagnostics.

Source & status

Source

Health Canada

Tier

beta

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

82270

Classification

Type II

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

HOME
Medium

Atellica ch enzymatic hemoglobin a1c (a1c e) assay

Health Canada18 Jun 2026

Brand match

Medium risk

Markets

CA

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 2 Jul 2026. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.