No image published for this recall

No image published for this recall

Ethibond excel® , ethibond®, permahand®, prolene® sutures

Official notice

9 Jan 2025

Last checked

9 Jan 2025

Medium riskHealth Canada

An official Health Canada notice concerns Ethibond excel® , ethibond®, permahand®, prolene® sutures. An open seal defect was detected during internal rework activities. Investigation determined that the preliminary cause of issue was a machine failure, leading to the zipper tray to be incorrectly positioned in the copolymer pre-formed cavity and subsequent tray misalignment with the front seal area in the sealing process. This ultimately resulted in possibility of open seal condition and/or narrow seal defects in the subsequent units. Recall start date: December 30, 2024 Published 2025-01-09.

What to do now

Contact the manufacturer if you require additional information.

Recall number

76812

Official notice

2025-01-09

The hazard

An open seal defect was detected during internal rework activities. Investigation determined that the preliminary cause of issue was a machine failure, leading to the zipper tray to be incorrectly positioned in the copolymer pre-formed cavity and subsequent tray misalignment with the front seal area in the sealing process. This ultimately resulted in possibility of open seal condition and/or narrow seal defects in the subsequent units. Recall start date: December 30, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

76812

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - ETHIBOND® Extra Polyester Suture, Lot or serial number - 103H4B, Model or catalog number - X425H

Product 2

Affected products - PROLENE® Suture - Taper Point, Lot or serial number - 103E22, Model or catalog number - 8706H

Product 3

Affected products - PERMAHAND® Non Absorbable Silk Suture, Lot or serial number - 10386S 103DBT, Model or catalog number - 623H K872H

Product 4

Affected products - ETHIBOND EXCEL® Polyester Suture, Lot or serial number - 103ELK, Model or catalog number - X872H

Risk / hazard

An open seal defect was detected during internal rework activities. Investigation determined that the preliminary cause of issue was a machine failure, leading to the zipper tray to be incorrectly positioned in the copolymer pre-formed cavity and subsequent tray misalignment with the front seal area in the sealing process. This ultimately resulted in possibility of open seal condition and/or narrow seal defects in the subsequent units. Recall start date: December 30, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Ethicon, Ethibond Excel® Ethibond® Permahand® Prolene® Sutures.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

76812

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 9 Jan 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.