Open official notice

Pds ii (polydiaxanone) absorbable suture

Official notice

18 Aug 2023

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Pds ii (polydiaxanone) absorbable suture. Ethicon received seven complaints for one lot of undyed PDS™ II sutures regarding low tensile strength. Internal testing on returned product from this lot confirmed that some PDS™ II sutures from this lot did not meet Ethicon's tensile strength requirement. An investigation identified additional lots across the PDS™ Plus Antibacterial and PDS™ II suture families that have the potential to be affected by the same root cause as the lot that received the complaints. Recall Start Date: August 8, 2023 Published 2023-08-18.

What to do now

Contact the manufacturer if you require additional information.

Recall number

74132

Official notice

2023-08-18

The hazard

Ethicon received seven complaints for one lot of undyed PDS™ II sutures regarding low tensile strength. Internal testing on returned product from this lot confirmed that some PDS™ II sutures from this lot did not meet Ethicon's tensile strength requirement. An investigation identified additional lots across the PDS™ Plus Antibacterial and PDS™ II suture families that have the potential to be affected by the same root cause as the lot that received the complaints. Recall Start Date: August 8, 2023

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

74132

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - PDS II (Polydiaxanone) Absorbable Suture, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - PDP442H PDP497G

Hazard

Ethicon received seven complaints for one lot of undyed PDS™ II sutures regarding low tensile strength. Internal testing on returned product from this lot confirmed that some PDS™ II sutures from this lot did not meet Ethicon's tensile strength requirement. An investigation identified additional lots across the PDS™ Plus Antibacterial and PDS™ II suture families that have the potential to be affected by the same root cause as the lot that received the complaints. Recall Start Date: August 8, 2023

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Ethicon, Pds Ii Absorbable Suture.

Source & status

Tier

official

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

74132

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 18 Aug 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button