No image published for this recall

No image published for this recall

Vitros® systems

Official notice

7 Mar 2024

Last checked

7 Mar 2024

Medium riskHealth Canada

An official Health Canada notice concerns Vitros® systems. Some serial numbers of the Vitros® systems as listed above may experience an anomaly involving the sampling center, which may have been manufactured using a sub-optimal hardware part. Quidelortho confirmed a complaint where the sampling center mechanism unintentionally disengaged from the universal sample tray (tray), resulting in the misalignment of the intended sample container to be aspirated. From our investigation of this single occurrence, Quidelortho determined that if this disengagement and misalignment occurs, it may cause one or multiple patient samples to be aspirated from an unintended sample container within the same tray and report the results using the incorrect patient samples. Based on the investigation, Quidelortho estimates the likelihood of this occurrence to be approximately 0.00000035% (or about 1 in 2.88 million tests.) Recall start date: January 26, 2024 Published 2024-03-07.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75187

Official notice

2024-03-07

The hazard

Some serial numbers of the Vitros® systems as listed above may experience an anomaly involving the sampling center, which may have been manufactured using a sub-optimal hardware part. Quidelortho confirmed a complaint where the sampling center mechanism unintentionally disengaged from the universal sample tray (tray), resulting in the misalignment of the intended sample container to be aspirated. From our investigation of this single occurrence, Quidelortho determined that if this disengagement and misalignment occurs, it may cause one or multiple patient samples to be aspirated from an unintended sample container within the same tray and report the results using the incorrect patient samples. Based on the investigation, Quidelortho estimates the likelihood of this occurrence to be approximately 0.00000035% (or about 1 in 2.88 million tests.) Recall start date: January 26, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75187

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Vitros 5600 Integrated System, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 680 2413 680 2915

Product 2

Affected products - Vitros Xt 7600 Integrated System, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 684 4461

Risk / hazard

Some serial numbers of the Vitros® systems as listed above may experience an anomaly involving the sampling center, which may have been manufactured using a sub-optimal hardware part. Quidelortho confirmed a complaint where the sampling center mechanism unintentionally disengaged from the universal sample tray (tray), resulting in the misalignment of the intended sample container to be aspirated. From our investigation of this single occurrence, Quidelortho determined that if this disengagement and misalignment occurs, it may cause one or multiple patient samples to be aspirated from an unintended sample container within the same tray and report the results using the incorrect patient samples. Based on the investigation, Quidelortho estimates the likelihood of this occurrence to be approximately 0.00000035% (or about 1 in 2.88 million tests.) Recall start date: January 26, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Ortho-clinical Diagnostics, Vitros®.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

75187

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 7 Mar 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.