Image not available in RecallScope.
Vitros® chemistry products system - lac slides (lactate)
Official notice
14 Aug 2025
An official Health Canada notice concerns Vitros® chemistry products system - lac slides (lactate). QuidelOrtho investigated customer complaints related to the inability to successfully calibrate VITROS® LAC slides on VITROS® XT 3400 systems and VITROS® XT 7600 systems. To resolve this issue, QuidelOrtho has adjusted the assay parameter used by the reflectometer to process VITROS® LAC slides. This parameter adjustment will be implemented beginning in add DRV 6339 (and above). QuidelOrtho advises affected customers to load add DRV 6339 (or above) upon receipt. Recall start date: August 5, 2025 Published 2025-08-14.
What to do now
Contact the manufacturer if you require additional information.
Recall number
77908
Official notice
2025-08-14
The hazard
The hazard
QuidelOrtho investigated customer complaints related to the inability to successfully calibrate VITROS® LAC slides on VITROS® XT 3400 systems and VITROS® XT 7600 systems. To resolve this issue, QuidelOrtho has adjusted the assay parameter used by the reflectometer to process VITROS® LAC slides. This parameter adjustment will be implemented beginning in add DRV 6339 (and above). QuidelOrtho advises affected customers to load add DRV 6339 (or above) upon receipt. Recall start date: August 5, 2025
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Product 2
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Manufacturer
Phone
Brand: Ortho-clinical Diagnostics, Vitros® Chemistry System - Lac Slides.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.