Vitros® 5600, 3600 and xt systems
Official notice
9 Jan 2025
An official Health Canada notice concerns Vitros® 5600, 3600 and xt systems. During internal testing, QuidelOrtho™ identified that microwell assay results have the potential to be reported without the reagent expired (re) result code, even when the signal reagent (sr) pack is past the expiry date. This can occur when an sr pack exceeds the on-board expiry and/or the shelf expiration date. This issue can affect any microwell assay regardless of whether or not the system is configured to use expired reagents. Recall start date: December 24, 2024 Published 2025-01-09.
What to do now
Contact the manufacturer if you require additional information.
Recall number
76811
Official notice
2025-01-09
The hazard
The hazard
During internal testing, QuidelOrtho™ identified that microwell assay results have the potential to be reported without the reagent expired (re) result code, even when the signal reagent (sr) pack is past the expiry date. This can occur when an sr pack exceeds the on-board expiry and/or the shelf expiration date. This issue can affect any microwell assay regardless of whether or not the system is configured to use expired reagents. Recall start date: December 24, 2024
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Product 2
Product 3
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Ortho-clinical Diagnostics, Vitros® 5600 3600 And Xt.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.