Vitros chemistry products system - crp slides (c-reactive protein)

Official notice

6 Nov 2024

Last checked

6 Nov 2024

Medium riskHealth Canada

An official Health Canada notice concerns Vitros chemistry products system - crp slides (c-reactive protein). QuidelOrtho™ received complaints on the affected coating of VITROS® CRP slides, related to lower-than-expected results when testing patient samples known to have elevated concentrations of CRP (>90 mg/l or >9 mg/dl). When tested neat, results were reported as being within the upper end of the measuring range (70-90 mg/l or 7-9 mg/dl) however, upon dilution, results for these samples were reported as having concentrations greater than 2x the neat result. Patient samples with a CRP concentration above 90 mg/l (9 mg/dl) (typically above 150 mg/l (15 mg/dl)) may be reported as falsely reduced when samples are tested neat, with the dilute sample results predicting greater than 2x greater than the neat result. The observed bias will vary; however, the predicted neat CRP results will generally range between 70-90 mg/l (7-9 mg/dl). Recall start date: October 25, 2024 Published 2024-11-06.

What to do now

Contact the manufacturer if you require additional information.

Recall number

76455

Official notice

2024-11-06

The hazard

QuidelOrtho™ received complaints on the affected coating of VITROS® CRP slides, related to lower-than-expected results when testing patient samples known to have elevated concentrations of CRP (>90 mg/l or >9 mg/dl). When tested neat, results were reported as being within the upper end of the measuring range (70-90 mg/l or 7-9 mg/dl) however, upon dilution, results for these samples were reported as having concentrations greater than 2x the neat result. Patient samples with a CRP concentration above 90 mg/l (9 mg/dl) (typically above 150 mg/l (15 mg/dl)) may be reported as falsely reduced when samples are tested neat, with the dilute sample results predicting greater than 2x greater than the neat result. The observed bias will vary; however, the predicted neat CRP results will generally range between 70-90 mg/l (7-9 mg/dl). Recall start date: October 25, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

76455

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Vitros Chemistry Products System - Crp Slides (C-Reactive Protein), Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 192 6740

Risk / hazard

QuidelOrtho™ received complaints on the affected coating of VITROS® CRP slides, related to lower-than-expected results when testing patient samples known to have elevated concentrations of CRP (>90 mg/l or >9 mg/dl). When tested neat, results were reported as being within the upper end of the measuring range (70-90 mg/l or 7-9 mg/dl) however, upon dilution, results for these samples were reported as having concentrations greater than 2x the neat result. Patient samples with a CRP concentration above 90 mg/l (9 mg/dl) (typically above 150 mg/l (15 mg/dl)) may be reported as falsely reduced when samples are tested neat, with the dilute sample results predicting greater than 2x greater than the neat result. The observed bias will vary; however, the predicted neat CRP results will generally range between 70-90 mg/l (7-9 mg/dl). Recall start date: October 25, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Ortho-clinical Diagnostics, Vitros Chemistry System - Crp Slides.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

76455

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 6 Nov 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.