Vitros immunodiagnostic products anti-hbs calibrators and reagent
Official notice
30 May 2024
An official Health Canada notice concerns Vitros immunodiagnostic products anti-hbs calibrators and reagent. QuidelOrtho's internal stability testing of VITROS anti-HBs calibrators detected a gradual reduction in calibration signal for calibrator levels 2 and 3 over the shelf life of the product for the specified lots. This reduction in signal may affect the shape of the calibration curve generated during calibration, compared to the calibration curve generated earlier in the product's shelf life. A change in the shape of the calibration curve may cause patient and QC samples tested using the associated VITROS® immunodiagnostic products anti-HBs reagent pack to report results as falsely elevated, potentially causing negative results to report as borderline, and borderline results to report as positive. This issue is detectable by performing QC as results generated will be higher than expected and/or outside the expected limits for QC, depending on the age of the lot and amount of signal reduction experienced in calibrator levels 2 and 3. Recall start date: May 21, 2024 Published 2024-05-30.
What to do now
Contact the manufacturer if you require additional information.
Recall number
75591
Official notice
2024-05-30
The hazard
The hazard
QuidelOrtho's internal stability testing of VITROS anti-HBs calibrators detected a gradual reduction in calibration signal for calibrator levels 2 and 3 over the shelf life of the product for the specified lots. This reduction in signal may affect the shape of the calibration curve generated during calibration, compared to the calibration curve generated earlier in the product's shelf life. A change in the shape of the calibration curve may cause patient and QC samples tested using the associated VITROS® immunodiagnostic products anti-HBs reagent pack to report results as falsely elevated, potentially causing negative results to report as borderline, and borderline results to report as positive. This issue is detectable by performing QC as results generated will be higher than expected and/or outside the expected limits for QC, depending on the age of the lot and amount of signal reduction experienced in calibrator levels 2 and 3. Recall start date: May 21, 2024
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Manufacturer
Phone
Brand: Ortho-clinical Diagnostics, Vitros Immunodiagnostic Anti-hbs Calibrators And Reagent.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.