Patient information centre ix

Official notice

20 Aug 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Patient information centre ix. Philips received reports that some PIC iX systems were frozen and could not be used. It was determined that failure of the watchdog may cause the system to be in an unknown state and therefore potentially prevent alarms from being generated or logged. Following the initial system error which triggers the watchdog, staff may observe unexpected system behavior while the watchdog attempts recovery. The observed behavior depends on the underlying software issue and may include:-Inability to interact with the PIC iX user interface. If real-time waveforms, numeric data and alarms remain operational, there is no impact to bedside or central monitoring and alarming.-The user will not recognize the device issue before the hazardous situation occurs if theobserved behavior is: Loss of central alarming, Loss of, or frozen, waveforms/data display-Bedside monitors and telemetry devices could generate a "No Central Mon" INOP Published 2026-08-20.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82511

Official notice

2026-08-20

The hazard

Philips received reports that some PIC iX systems were frozen and could not be used. It was determined that failure of the watchdog may cause the system to be in an unknown state and therefore potentially prevent alarms from being generated or logged. Following the initial system error which triggers the watchdog, staff may observe unexpected system behavior while the watchdog attempts recovery. The observed behavior depends on the underlying software issue and may include:-Inability to interact with the PIC iX user interface. If real-time waveforms, numeric data and alarms remain operational, there is no impact to bedside or central monitoring and alarming.-The user will not recognize the device issue before the hazardous situation occurs if theobserved behavior is: Loss of central alarming, Loss of, or frozen, waveforms/data display-Bedside monitors and telemetry devices could generate a "No Central Mon" INOP

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

RA RA-82511

Recall Numbers

82511

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - Not applicable., Model or catalogue number - VERSION 4.2

Product 2

Lot or serial number - Not applicable., Model or catalogue number - VERSION 4.0

Product 3

Lot or serial number - Not applicable., Model or catalogue number - 4.5

Product 4

Lot or serial number - Not applicable., Model or catalogue number - VERSION 4.3

Product 5

Lot or serial number - Not applicable., Model or catalogue number - 4.4

Product 6

Lot or serial number - Not applicable., Model or catalogue number - VERSION 4.1

Hazard

Philips received reports that some PIC iX systems were frozen and could not be used. It was determined that failure of the watchdog may cause the system to be in an unknown state and therefore potentially prevent alarms from being generated or logged. Following the initial system error which triggers the watchdog, staff may observe unexpected system behavior while the watchdog attempts recovery. The observed behavior depends on the underlying software issue and may include:-Inability to interact with the PIC iX user interface. If real-time waveforms, numeric data and alarms remain operational, there is no impact to bedside or central monitoring and alarming.-The user will not recognize the device issue before the hazardous situation occurs if theobserved behavior is: Loss of central alarming, Loss of, or frozen, waveforms/data display-Bedside monitors and telemetry devices could generate a "No Central Mon" INOP

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Philips Medizin Systeme Boblingen.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82511

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 20 Aug 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button