An official Health Canada notice concerns Intellispace critical care and anesthesia. In ICCA releases J.06.01 and J.06.02, calculations triggered by infusion-pump auto-charting may result in incorrect Intake Totals, Output Totals, Net Body Balance, Length of Stay (LOS) Balance, and related fluid-balance values.This issue applies only to customers that have configured and use Auto Charting for Infusion Data for a Clinical Unit, with one or both of the following settings enabled:• Automatic periodic charting of infusion rates and volumes.• Automatic aperiodic charting of infusion rates and bolus volumes when a change occurs.Customers that do not have either Auto Charting for Infusion Data setting configured and in use are not impacted by this issue. Published 2026-09-16.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82635

Official notice

2026-09-16

The hazard

In ICCA releases J.06.01 and J.06.02, calculations triggered by infusion-pump auto-charting may result in incorrect Intake Totals, Output Totals, Net Body Balance, Length of Stay (LOS) Balance, and related fluid-balance values.This issue applies only to customers that have configured and use Auto Charting for Infusion Data for a Clinical Unit, with one or both of the following settings enabled:• Automatic periodic charting of infusion rates and volumes.• Automatic aperiodic charting of infusion rates and bolus volumes when a change occurs.Customers that do not have either Auto Charting for Infusion Data setting configured and in use are not impacted by this issue.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

RA RA-82635

Recall Numbers

82635

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - Not applicable., Model or catalogue number - VERSION J.06

Hazard

In ICCA releases J.06.01 and J.06.02, calculations triggered by infusion-pump auto-charting may result in incorrect Intake Totals, Output Totals, Net Body Balance, Length of Stay (LOS) Balance, and related fluid-balance values.This issue applies only to customers that have configured and use Auto Charting for Infusion Data for a Clinical Unit, with one or both of the following settings enabled:• Automatic periodic charting of infusion rates and volumes.• Automatic aperiodic charting of infusion rates and bolus volumes when a change occurs.Customers that do not have either Auto Charting for Infusion Data setting configured and in use are not impacted by this issue.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Philips Medizin Systeme Boblingen.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82635

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 16 Sept 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button