An official Health Canada notice concerns Siemens healthineers ag magnetom and biograph one systems. A software issue has been identified in syngo.via MR Cardiac Analysis MR Flow Quantification affecting the evaluation of retrospectively gated 2D cardiac flow data. The algorithm does not correctly represent the final phase of the cardiac cycle. The evaluation stops at the last trigger time point (TT point), which marks the beginning of the final cardiac phase, instead of accounting for the complete duration of that phase until the end of the cardiac cycle. The final repetition-time interval (TR) is not included in the calculation (“TR missing”). This results in systematic truncation of the cardiac cycle and incomplete loop closure. Consequently, flow quantification parameters (e.g., Forward volume, Avg.Flow/Min) are systematically underestimated. The impact is relevant for Beat Sensor-triggered flow data, for which the flow peak occurs in the very last acquired cardiac phases. Published 2026-10-28.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82872

Official notice

2026-10-28

The hazard

A software issue has been identified in syngo.via MR Cardiac Analysis MR Flow Quantification affecting the evaluation of retrospectively gated 2D cardiac flow data. The algorithm does not correctly represent the final phase of the cardiac cycle. The evaluation stops at the last trigger time point (TT point), which marks the beginning of the final cardiac phase, instead of accounting for the complete duration of that phase until the end of the cardiac cycle. The final repetition-time interval (TR) is not included in the calculation (“TR missing”). This results in systematic truncation of the cardiac cycle and incomplete loop closure. Consequently, flow quantification parameters (e.g., Forward volume, Avg.Flow/Min) are systematically underestimated. The impact is relevant for Beat Sensor-triggered flow data, for which the flow peak occurs in the very last acquired cardiac phases.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

234786, 234723, 234827, 234832, 196680, 238015, 238023, 202203, 202080, 239040

Skus

11291455, 11410482, 11851911, 11344916, 11689172, 11410481, 11060815, 11689652

Other

RA RA-82872

Model numbers

11291455, 11410482, 11851911, 11344916, 11689172, 11410481, 11060815, 11689652

Recall Numbers

82872

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 11291455

Product 2

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 11410482

Product 3

Lot or serial number - 234786, Model or catalogue number - 11851911

Product 4

Lot or serial number - 234723, Model or catalogue number - 11851911

Product 5

Lot or serial number - 234827, Model or catalogue number - 11851911

Product 6

Lot or serial number - 234832, Model or catalogue number - 11851911

Product 7

Lot or serial number - 196680, Model or catalogue number - 11344916

Product 8

Lot or serial number - 238015, Model or catalogue number - 11689172

Product 9

Lot or serial number - 238023, Model or catalogue number - 11689172

Product 10

Lot or serial number - 202203, Model or catalogue number - 11410481

Product 11

Lot or serial number - 202080, Model or catalogue number - 11410481

Product 12

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 11060815

Product 13

Lot or serial number - 239040, Model or catalogue number - 11689652

Hazard

A software issue has been identified in syngo.via MR Cardiac Analysis MR Flow Quantification affecting the evaluation of retrospectively gated 2D cardiac flow data. The algorithm does not correctly represent the final phase of the cardiac cycle. The evaluation stops at the last trigger time point (TT point), which marks the beginning of the final cardiac phase, instead of accounting for the complete duration of that phase until the end of the cardiac cycle. The final repetition-time interval (TR) is not included in the calculation (“TR missing”). This results in systematic truncation of the cardiac cycle and incomplete loop closure. Consequently, flow quantification parameters (e.g., Forward volume, Avg.Flow/Min) are systematically underestimated. The impact is relevant for Beat Sensor-triggered flow data, for which the flow peak occurs in the very last acquired cardiac phases.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Siemens Healthineers.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82872

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 28 Oct 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button